DSX1000
While pushing the zoom head to the microscope frame, rotate the zoom head counterclockwise until it stops, and
tighten the clamping screws (3 positions)
provided with the zoom head using the Allen wrench provided with the
zoom head.
Insert the tip of the long side of Allen wrench into the clamping screw and tighten it temporarily. And then, insert the
tip of the short side of Allen wrench into the clamping screw and tighten it completely.
It is easier to tighten from the upper screw.
Detach the transport lock screw using the Allen screwdriver.
NOTE
l
If you move the zoom head without detaching the transport screw, the product could be damaged. Be
sure to detach it.
l
As the transport screw will be used for transporting the system next time, keep it in a safe place.
Detaching the zoom head
Move the zoom head until the Z position is between -48,000 to -49,000. For details, see "3.2 Vertical movement of the
zoom head" in Operation manual.
When the tilting frame DSX10-TF is combined, return the zoom head to the upright position. For the operation
method, see "3.6 Tilting observation" in Operation manual.
Set not to initialize the zoom head position when the application is closed. For details, see "15.1 Initializing the zoom
head when closing the application" in Operation manual.
Attach the transport lock screw using the Allen screwdriver. For details, see
"Attachment/detachment of the zoom
32
8. Assembly
Downloaded from