Problem
Cause
Remedy
Page
The picture is not in focus.
The specimen is not in focus
properly.
Adjust the focus correctly with the
fine adjustment knob.
–
The aperture iris diaphragm of the
condenser is open too wide.
Close the aperture iris diaphragm
a little.
–
The field iris diaphragm is not set
properly.
Adjust the field iris diaphragm until
the image circumscribes the field
of view.
–
Lens components of the micro-
scope are contaminated or the
cover glass on the front of the
camera is stained.
Clean the objective, camera
adapter lens, condenser and/or
window lens of the microscope,
or clean the cover glass on the
bottom of the camera head.
14
The microscope and/or camera
are subjected to vibration during
recording.
Acquire images in an environment
in which the microscope and
camera are not vibrated. It is effec-
tive to use an anti-vibration bench.
–
The 4140 x 3096 and 2070 x 1548
images are not neat.
The camera is subjected to
vibration during recording.
Acquire images in an environment
in which the microscope and
camera are not vibrated. It is effec-
tive to use an anti-vibration bench.
–