28
Problem
Cause
Remedy
Page
The picture is not in focus.
The microscope is not focused
properly.
Adjust the focus correctly with the fine
adjustment knob.
–
The parfocality is not adjusted
properly.
Adjust parfocality with the camera
adapter.
–
The aperture iris diaphragm of the
condenser is open too wide.
Close the aperture iris diaphragm a
little.
–
The field iris diaphragm is not set
properly.
Adjust the field iris diaphragm until the
image circumscribes the field of view.
–
Lens components of the microscope
are contaminated or the cover glass
on the front of the camera is stained.
Clean the objective, photography lens,
condenser and/or window lens of the
microscope, or clean the cover glass
on the bottom of the camera head.
6
The microscope and/or camera
are subjected to vibration during
recording.
Record images in an environment in
which the microscope and camera
are not vibrated. It is effective to use an
anti-vibration bench.
–
The DP2-TWAIN or cellSens /
OLYMPUS Stream window is not
displayed correctly or the menu
characters are not displayed correctly.
The resolution setting of the screen is
incorrect.
Set the resolution setting at 1280 x
1024 or more in the properties of the
screen.
–
The large font has been selected for
the font size of the screen.
Select a small font in the properties of
the screen.
–
Another 1394 device cannot be used
simultaneously.
The DP26-dedicated driver is installed. When the DP26 is connected to a PC,
other 1394 devices cannot be used on
the same PC.
10,21
DP26