background image

103076-01EN [U8998318], Rev. C, June 2014

8    Safety Information

   

 

When shipping the Li-ion battery, be sure to follow all local transportation regulations.

3. Safety Information

3.1

Radiation Safety Information

Olympus handheld XRF analyzers are secure and dependable instruments when used in accordance with recommended 
testing techniques and safety procedures.

Olympus analyzers must only be used by trained and authorized operators in accordance with proper safety 
procedures. Improper use may impair safety protection and cause potential harm to the user. 

Read all warning signs and labels.

DO NOT USE the instrument if there is any evidence of damage, as doing so could cause unintentional emission of 
stray radiation. In such a case, arrange for qualified personnel to perform a radiation safety test and repair any 
damage to the instrument.

3.2

Safety Interlock Structure

To control X-ray emissions, thereby minimizing the possibility of accidental exposure, the DELTA handheld XRF analyzer 
has a standard safety interlock structure consisting of the three features listed below:

1.

Software proximity sensor 

Within one second of a test start, the analyzer detects the sample in front of the measurement window. If no sample is 
detected the test aborts to prevent accidental exposure, the filter wheel goes to position 0, and the X-rays shut off. The 
tube current is reduced to 0.0

μ

A, and the red light stops blinking. Also, if the probe/nose is pulled away from the 

sample while a test is in progress, testing stops in approximately one second.

2.

Software trigger lock

 

If five minutes elapse between tests (default time), the trigger automatically locks and you must tap on the Lock button 

(

) to unlock it.

3.

Safeguards

As an owner of an Olympus handheld XRF analyzer, your safeguards include those items indicated below:

Limited access
Keep the instrument in a controlled location to which only trained and authorized users have access.

Trained operators
Post a sign near the analyzer indicating that it must only be used by operators who have completed a training 
class provided by your company, or who have attended an Olympus training course and completed any other 
requirements stipulated by the local regulating authority. When the Olympus instrument is turned on, the user 
interface screen displays a message indicating that the instrument should only be used by authorized personnel.

Shielding  issues
An Olympus handheld XRF analyzer emits a tightly collimated beam of X-ray radiation. Although attenuation 
occurs, the beam may project many meters in open air.
Adequate shielding is achieved by:

— Establishing a no-admittance zone sufficiently distant from the instrument’s measurement window to allow

air to attenuate the beam.

Содержание DELTA 50

Страница 1: ...us product safely and effectively Before using this product thoroughly review this instruction manual Use the product as instructed Keep this instruction manual in a safe accessible location Models DE...

Страница 2: ...e instrument 1 Locate and remove the shipping papers and documentation 2 Open the carrying case and then remove the DELTA and all of the components Please note that there are two layers of protective...

Страница 3: ...omponent Key DELTA All models except DELTA Element and DELTA Inspector Foam top layer 1 Carry case 2 DELTA analyzer 3 Docking station charger 4 USB cable 1 see page 6 for details 5 USB cable 2 see pag...

Страница 4: ...del shown 2 Probe 3 Measurement window Prolene or Kapton film 4 Hinged window plate 5 Docking station connector 6 Trigger 7 Handle Non slip rubber grip 8 Battery boot 9 Data port with rubber cover 10...

Страница 5: ...DELTA Element and DELTA Inspector models This battery charger is described in section 4 2 2 on page 11 Component Key DELTA All models except DELTA Element and DELTA inspector Docking station empty 1 A...

Страница 6: ...omponent key DELTA All models standard on the DELTA Element and DELTA Inspector Battery charger empty 1 Input power 12 VDC 2 Battery cradle and connectors 3 Charging error indicator 4 Battery charging...

Страница 7: ...and either Secure it with plastic zip ties or Pack the transport case within another box USB cable 2 Mini USB B to USB A connectors AC power adaptor Provides DC power to docking station Input 110 240...

Страница 8: ...accidental exposure the filter wheel goes to position 0 and the X rays shut off The tube current is reduced to 0 0 A and the red light stops blinking Also if the probe nose is pulled away from the sam...

Страница 9: ...e I O switch to turn on the power A green LED indicator illuminates This switch DOES NOT turn the X ray tube on No tube power is supplied until the Olympus software is launched POWER OFF Press and hol...

Страница 10: ...percent of charge remaining ranging from less than 25 to 100 If a battery has a charge of less than 25 use the docking station to perform a full charge Figure 4 1 Battery charge status button 4 2 Char...

Страница 11: ...y A battery hot swap capability is a standard feature with the DELTA handheld XRF analyzer You can remove and replace a battery without having to shut down restart or perform a Cal Check A shutdown st...

Страница 12: ...ld the analyzer upside down by the handle with the bottom of the analyzer base pointing upward and the nose pointing away from the operator 2 Pull the rubber latch and lift the cover 3 Pull the tab to...

Страница 13: ...up radiation safety and initialization screens see chapter Operation on page 15 The main operations are accessible from the Home screen see Figure 5 1 on page 13 Refer to the User Interface Guide P N...

Страница 14: ...y Spectrum Display Chemistry Exit screen and save changes Exit screen and don t save changes Display Expand Display Minimize X Review S W Title Name of screen and current mode Padlock trigger indicato...

Страница 15: ...safety notice see and then confirm that you are a certified user System initialization begins immediately The analyzer launches a test screen using the last selected mode see Figure 6 1 on page 15 4 I...

Страница 16: ...6 3 on page 16 Figure 6 3 Results 11 At the end of the session export the results to a PC using the data port and USB cable 12 When testing and exporting are complete turn off the analyzer using the...

Страница 17: ...ter confirmation 3 Relaunch Restart the application All user interface UI screens have a time out power saving feature that causes the screen to go blank if the UI is not accessed or if the instrument...

Страница 18: ...103076 01EN U8998318 Rev C June 2014 18 Operation...

Страница 19: ...2014 DELTA Radiation Profile 19 7 DELTA Radiation Profile This chapter includes the radiation profiles for the DELTA handheld XRF analyzers Figure 7 1 DELTA radiation profile Handheld instruments 61 c...

Страница 20: ...4 Beam RoHS 2 alloy PVC Blank 50e e All 50 kV readings are taken with standard probe shield in place 80 Copper 10 840 400 250 15 EC 680K 16 1800 800 500 30 71X SR2 solder BK 5 1 BK BK Alloy Mining 31...

Страница 21: ...200 120 60 22 7 EC 680K 60 9 1500 1000 1500 1000 750 200 56 37 71X SR2 solder BK BK BK BK BK BK BK BK BK BK Al 319 AA 1 BK 25 15 10 10 6 5 BK BK Alloy Mining 316 stainless 35 100 Aluminum BK BK BK BK...

Страница 22: ......

Страница 23: ...oltage warning symbol This symbol is used to alert the user to potential electric shock hazards greater than 1000 volts All safety messages that follow this symbol shall be obeyed to avoid possible ha...

Страница 24: ...otify the carrier making the delivery of any damage because the carrier is normally liable for damage during shipment Retain packing materials waybills and other shipping documentation needed in order...

Страница 25: ...103076 01EN U8998318 Rev C June 2014 Important Information Please Read Before Use 27...

Страница 26: ...MA 02453 USA www olympus ims com Printed in the United States of America Copyright 2011 2013 2014 by Olympus All rights reserved All brands are trademarks or registered trademarks of their respective...

Отзывы: