background image

15

CX23

Objectives

Magnifi-

cation

Numerical

aperture

N.A.

Numerical aperture of condenser suitable for

 observation of microscope specimens

(when setting to 70% of numerical aperture of objective lens)

Plan objective lens
(FN 20)

CX22PL100XO (option)

4X

10X

40X

100XO

0.10

0.25
0.65

1.25

0.1 X 0.7 = 0.07

0.25 X 0.7 = 0.175

0.65 X 0.7 = 0.455

1.25 X 0.7 = 0.875

Aperture stop

The aperture stop of the condenser is a mechanism to adjust the numerical aperture of the condenser to make the incidence 
angle of the illumination light to the value suitable for observation.

The observation is available with optimum contrast and resolution by adjusting the numerical aperture of the condenser 
appropriately for the numerical aperture of each objective lens.

In general, since the contrast of the microscope specimen is low, it is appropriate to set the numerical aperture of the 
condenser to approximately 70 to 80% of the numerical aperture of the objective lens.

The condenser of this product is designed so that the numerical aperture of the condenser is approximately 70% of the 
numerical aperture of the objective lens by coinciding the aperture stop lever position with the magnification display position 
corresponding to the magnification of the objective lens to be used.

Numerical aperture of objective lens and numerical aperture of condenser suitable for observation

Содержание CX23

Страница 1: ...ty obtain optimum performance and to familiarize yourself fully with the use of this microscope we recommend that you study this manual thoroughly before operating the microscope Retain this instructi...

Страница 2: ...re designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment This equipment generates uses and can radiate radio frequency energ...

Страница 3: ...3 Handling precautions 3 Maintenance and storage 4 1 Standard combination 5 2 Nomenclature of operating portions 6 3 Summary of brightfield observation procedure 7 4 Detailed observation procedure 8 1...

Страница 4: ...cs 19 9 Assembly 20 10 Optional accessories 21 10 1 System diagram of optional accessories 21 10 2 Installation and operation of optional accessories 21 1 Filter holder CH2 FH and darkfield ring CH2 D...

Страница 5: ...such as gloves etc to prevent the specimens from touching your skin directly When you maintain the device which may have contacted the specimens which have potentiality of infection wear protective eq...

Страница 6: ...at of the power outlet If the device is not grounded our intended electric safety and EMC performance of the device can not be assured Do not use the device in close proximity to the sources of strong...

Страница 7: ...or observation tube b since they may be damaged Also when you carry the microscope remove specimens or filters in advance Otherwise they may be dropped When carrying the microscope be careful not to...

Страница 8: ...se organic solvents to clean the parts other than the glass components since it deteriorates painted surfaces or plastic parts 3 When not using the product store it in a dry place or cover it with a d...

Страница 9: ...roscope frame Power cord 4X 10X 40X 100X CX23LEDRFS1 4X 10X 40X CX23LEDRFS2 Binocular observation tube Eyepieces Fixed at 10X Revolving nosepiece Condenser Stage Objective 100X Wooden case CH20 WB2 Ey...

Страница 10: ...chapter 9 Assembly on page 20 For detailed explanation of each operation portion refer to the page indicated inside parentheses Cautions when unpacking The items 1 to 3 described below have been attac...

Страница 11: ...brightness 5 Revolving nosepiece p 12 6 Coarse fine focus adjustment knobs p 9 2 Brightness adjustment knob p 8 9 Aperture stop lever p 10 5 Revolving nosepiece p 12 6 Coarse fine adjustment knobs p 9...

Страница 12: ...in the direction of the arrow set the specimen by sliding it on the stage from the front toward the rear After setting the specimen return the specimen holding lever c gently Rotating the upper Y axi...

Страница 13: ...should be read at the position a on the specimen holder The Y axis scale should be read at the position of index line b 1 2 Working distance WD The WD means the distance between the objective and the...

Страница 14: ...nd to reduce eye strain during observation While keeping the right and left eyepieces horizontally move the binocular portion in the a or b direction to coincide the right view field with the left com...

Страница 15: ...nd confirm that the right and left eyepieces are focused on the specimen If the specimen is not focused bring the specimen into focus according to the procedure in No 6 and repeat No 7 to 9 When the e...

Страница 16: ...by holding the objective lens CAUTION 7 Switching the objectives Using the eye shades When wearing eyeglasses Use the eye shades in the normal folded down position When not wearing eyeglasses Extend t...

Страница 17: ...rsion objective by one reciprocation or two The condenser of this microscope provides the full performance when oil is placed between the slide glass and the front lens of the condenser If oil is not...

Страница 18: ...with the combinations which have higher magnification of objective The field number is the diameter of the image observed through an eyepiece indicated in millimeters Eyepiece 10X 20 mm WHSZ15X H 16 m...

Страница 19: ...lable with optimum contrast and resolution by adjusting the numerical aperture of the condenser appropriately for the numerical aperture of each objective lens In general since the contrast of the mic...

Страница 20: ...s not engaged in the light path Engage the objective in the light path by rotating the revolving nosepiece until its clicks 12 The eyepiece condenser window lens and or specimen slide are dirty Clean...

Страница 21: ...eyes are not compensated Compensate it correctly 11 Different eyepieces are used between the left side and right side Use same eyepiece on the left side and right side 22 12 Objective hits the specim...

Страница 22: ...ge Size 120 x 132 mm with mechanical stage Movement range X direction 76 mm x Y direction 30 mm Specimen holder Holds a single specimen Condenser Type Abbe condenser N A 1 25 when immersed in oil Aper...

Страница 23: ...r glass and the top of the objective when in focus Number of aperture NA This value corresponds to the F number of the camera and is related to the resolution Resolution becomes higher when the NA bec...

Страница 24: ...Connect the power cord s connector d to connector e firmly Connect the power cord s plug f to a wall power outlet g CAUTION 1 CAUTION 2 3 f g a b c e d 2 Storing the AC adapter and power cord Make sur...

Страница 25: ...field ring Insert the darkfield ring a in the CH2 FH filter holder b Push the filter holder b containing the darkfield ring a into the bottom of the condenser until it clicks into place The darkfield...

Страница 26: ...ece to which the eyepiece micrometer will be attached Hold the eyepiece frame and loosen the reticle holder a to remove it Place the micrometer disk a into the reticle holder b with the indications fa...

Страница 27: ...rror to get brighter field of view A planar reflection mirror is used normally However if the brightness of observed field of view is uneven use a concave reflection mirror CAUTION 1 2 a If the wooden...

Страница 28: ...ppliance coupling Table 1 Certified cord A power supply cord should be certified by one of the agencies listed in Table 1 or comprised of cordage marked with an agency marking per Table 1 or marked pe...

Страница 29: ...ic Cables BASEC BASEC HAR 10 10 30 N V KEMA KEMA KEUR HAR 10 30 30 SEMKO AB Svenska Elektriska Materielkontrollanstalter SEMKO HAR 10 10 50 sterreichischer Verband f r Elektrotechnik VE VE HAR 30 10 5...

Страница 30: ...MEMO...

Страница 31: ......

Страница 32: ...Olympus Tower 446 Bongeunsa ro Gangnam gu Seoul Korea 135 509 48 Woerd Avenue Waltham MA 02453 U S A Distributed by Shinjuku Monolith 2 3 1 Nishi Shinjuku Shinjuku ku Tokyo 163 0914 Japan Manufacture...

Отзывы: