[Symptom: Brightness is insufficient.]
TROUBLESHOOTING
Start
Check the lamp for
deterioration (darkness).
Measure the voltage applied to
the lamp (with the lamp
connected).
100 V class 15 V±0.75 V
200 V class 14.6 V±0.75 V
Replace the lamp.
Mount the lamp correctly.
Check the supply voltage.
Connect power that
meets the rating.
4-5
Is the lamp normal?
Is the lamp correctly
mounted?
Does the applied
voltage meet the
rating?
Is the mechanical iris
normal?
Does the supply
voltage meet the
rating?
The transformer is defective.
Replace the lamp.
Repair the mechanical iris.
NO
NO
NO
NO
NO
Содержание CLH-SC
Страница 1: ...MAINTENANCE MANUAL CLH SC...
Страница 10: ...2 2 DESCRIPTION OF BLOCK UNIT NAME BLOCK DIAGRAM MODEL CLH SC...
Страница 11: ...3 CIRCUIT DIAGRAM CIRCUIT DIAGRAM 3 1 UNIT NAME 120V DATE 99 01 MODEL CLH SC PAGE 1 4 120V...
Страница 12: ...3 2 220V CIRCUIT DIAGRAM UNIT NAME 220V DATE 99 01 MODEL CLH SC PAGE 2 4...
Страница 13: ...CIRCUIT DIAGRAM 3 3 230V UNIT NAME 230V DATE 99 01 MODEL CLH SC PAGE 3 4...
Страница 14: ...CIRCUIT DIAGRAM 3 4 240V UNIT NAME 240V DATE 99 01 MODEL CLH SC PAGE 4 4...
Страница 26: ...6 1 COVER SHASSY CLH SC FIG 1 3 6 EXPLODED PARTS DIAGRAM EXPLODED PARTS DIAGRAM...
Страница 27: ...6 2 CLH SC MAIN UNIT FIG 2 3 EXPLODED PARTS DIAGRAM...
Страница 28: ...6 3 CLH SC CABLE FIG 3 3 EXPLODED PARTS DIAGRAM...