![Olympus CLH-SC Скачать руководство пользователя страница 16](http://html1.mh-extra.com/html/olympus/clh-sc/clh-sc_maintenance-manual_3915066016.webp)
TROUBLESHOOTING
[Symptom: Power is not supplied.]
The supply voltage is abnormal,
or the power cord is defective.
The interlock switch is
defective.
The power switch is defective.
The thermo-fuse is broken.
The transformer is defective.
4-2
Start
Is power provided at
the inlet?
Is power provided in the
input side of the power
switch?
Is power provided in the
output side of the power
switch?
Is the thermo-fuse
conducting?
Is the voltage at the
terminals E1 and E3 of
UPCH10U 15 V AC?
Check conduction of the wires and
connection of the terminals and
connectors.
End
NO
NO
NO
NO
NO
Содержание CLH-SC
Страница 1: ...MAINTENANCE MANUAL CLH SC...
Страница 10: ...2 2 DESCRIPTION OF BLOCK UNIT NAME BLOCK DIAGRAM MODEL CLH SC...
Страница 11: ...3 CIRCUIT DIAGRAM CIRCUIT DIAGRAM 3 1 UNIT NAME 120V DATE 99 01 MODEL CLH SC PAGE 1 4 120V...
Страница 12: ...3 2 220V CIRCUIT DIAGRAM UNIT NAME 220V DATE 99 01 MODEL CLH SC PAGE 2 4...
Страница 13: ...CIRCUIT DIAGRAM 3 3 230V UNIT NAME 230V DATE 99 01 MODEL CLH SC PAGE 3 4...
Страница 14: ...CIRCUIT DIAGRAM 3 4 240V UNIT NAME 240V DATE 99 01 MODEL CLH SC PAGE 4 4...
Страница 26: ...6 1 COVER SHASSY CLH SC FIG 1 3 6 EXPLODED PARTS DIAGRAM EXPLODED PARTS DIAGRAM...
Страница 27: ...6 2 CLH SC MAIN UNIT FIG 2 3 EXPLODED PARTS DIAGRAM...
Страница 28: ...6 3 CLH SC CABLE FIG 3 3 EXPLODED PARTS DIAGRAM...