DMTA069-01EN, Rev. D, March 2017
Preparing the ChainSCANNER for an Inspection
51
8.
Adjust the probe or transducer position (see “To adjust the distance between the
probes” on page 31).
9.
Connect the appropriate cables and tubes to the probes and wedge.
10. Pass the probe cables and irrigation tubes under the module’s black ergonomic
cover (see Figure 3-2 on page 51).
Failure to pass the cables and tubes underneath the module’s ergonomic cover can
result in lack of pressure between the wedge and the part being inspected.
Figure 3-2 Cables passing underneath the module’s ergonomic cover
(model equipped with Four-Probe kit shown)
11. Cover the cables and tubes with the supplied divisible cable protection.
IMPORTANT
Scanner cables
passing underneath
cover
Ergonomic cover
Umbilical sheath
strain relief assembly
Divisible cable protection