background image

Phenomena

Cause

Remedy

Reference

page

Wrong cable is used.

Use a correct cable.

23

Wrong command is used.

Refer to the command reference

manual.

The nosepiece does not move.

The nosepiece setting to select

either the 6Position-NP or

5Position-NP on the DIP switch is

wrong.

Specify the correct nosepiece

setting on the DIP switch.

10

Only a specific hole (partial

segment) is used.

Refer to “To extend the life of the

nosepiece” and evenly apply the

grease.

31

Auto-focusing does not work.

The safety lock is not attached.

Attach the short plug provided

with the product or a lock

mechanism.

Repair request

If you cannot improve the phenomena after taking the above remedies, please contact us for assistance.

At that time, please tell them the following information as well.

l

Product name and abbreviation (Example: BXC-CBB)

l

Product number

l

Phenomena

34

6.3 Drive system

6. Troubleshooting

Содержание BXC-CBB

Страница 1: ...ety obtain optimum performance and to familiarize yourself fully with the use of this product we recommend that you study this manual thoroughly before operating this product and always keep this manu...

Страница 2: ...posed of as unsorted municipal waste but should be collected separately Refer to your local our distributor in EU for return and or collection systems available in your country Research and Industrial...

Страница 3: ...e light source for reflected light illumination 14 3 10 1 Attaching the LED illumination 14 3 10 2 Attaching the liquid light guide or mercury lamp housing 15 3 10 3 Attaching the motorized reflected...

Страница 4: ...5 Maintenance and Storage 31 5 1 Cleaning of each part 31 6 Troubleshooting 32 6 1 Optical systems 32 6 2 Observation tube 33 6 3 Drive system 33 7 Proper selection of the power supply cord 35 ii...

Страница 5: ...XC CBB Application manual Advanced BXC FSU Usage For details contact us BXC CBB Command reference manual How to use RS 232C communication commands For details contact us Intended use This product is d...

Страница 6: ...ing label position Right side surface of BXC FSU This product is specified as CLASS 1 laser product CLASS 1 LASER PRODUCT IEC60825 1 2007 IEC60825 1 2014 This product complies with 21 CFR 1040 10 and...

Страница 7: ...case of emergency disconnect the power cord from the power cord connector on the product or from the power outlet Install the product at the location where you can reach the power cord connector or t...

Страница 8: ...ectric safety Caution for prevention of electric shock 3 2 If a caution engraving or label is dirty or peeled off contact us for the replacement or other inquiry 2 8 Specifications Rating AC adapter I...

Страница 9: ...2 9 Restrictions The maximum total length of the cables BXC LCBL1M BXC LCBL3M BXC LCBL6M that can be used is 13 m 5 2 Safety precautions 2 9 Restrictions...

Страница 10: ...BXC RLI LGCA facing down on a desk and remove the transport lock NOTE l A nosepiece cannot be attached unless the transport lock is removed l If you start the device with the transport lock attached...

Страница 11: ...BXC RLI LGCA according to the following steps The following steps are described using only BXC RLI Tool used Allen wrench 3mm NOTE l Be sure to adjust centering of the Autofocus Sensing Unit BXC FSU...

Страница 12: ...mounting seat of the Autofocus Sensing Unit BXC FSU and insert the centering telescope CT D in the standard cylindrical shaft Rotate the helicoid E on the top of the CT to focus on the cross scale 3...

Страница 13: ...the CT then cover four holes E on top of the Autofocus Sensing Unit BXC FSU with the provided seals 3 4 Installing the control box When securing the control box to the floor wall or ceiling attach it...

Страница 14: ...e 1 2 3 4 5 6 7 8 Off Buzzer sound The buzzer sound is heard On The buzzer sound is not heard Off Nosepiece NP 5Position NP On 6Position NP Off AS control 1 Normal On High precision control lost motio...

Страница 15: ...sary High precision control By selecting high precision control you can reduce the effect of the lost motion In this mode the positioning accuracy improves as the system always stops at the specified...

Страница 16: ...the highest magnification objective l Attach the objectives starting from the nosepiece mounting hole 1 without leaving any holes in between empty l Make sure to place caps on any holes to which objec...

Страница 17: ...or cause poor connection of the connector 3 While pushing the nosepiece in the attaching direction with your left hand hold the Allen screwdriver between the pointing finger and the thumb and tighten...

Страница 18: ...sing an Allen screwdriver NOTE For safety turn OFF the powers of the control box and PC while connecting cables 2 Insert the reflected LED light source into the light source mounting hole of the refle...

Страница 19: ...l it touches the end and tighten two mounting screws A 3 10 3 Attaching the motorized reflected filter wheel U FWR 1 Loosen two mounting screws A of the reflected light illuminator using an Allen scre...

Страница 20: ...d LED light source or the halogen lamp housing into the motorized reflected filter wheel until it touches the end and tighten two mounting screws B 16 3 10 Attaching the light source for reflected lig...

Страница 21: ...e double lamp housing adapter U DULHA connect BX3M LEDR to BXC RLI using the extension cable for LED light source MX LLHECBL sold separately 3 10 5 Attaching the light source for motorized light guide...

Страница 22: ...tems you will use complies with standards required in your local area 3 11 Setting up BXC FSU 3 11 1 Product outline of BXC FSU BXC FSU is an intermediate mounting unit that is mounted on the top of t...

Страница 23: ...f BXC CBE1 1 BXC CBB I F Label CBB Description Internal I F power supply and communication signals 2 Focus Error signal Label E SIG Description Focus Error Signal Analog signal Purpose Used for Z axis...

Страница 24: ...S LOCK Description In Capture and In Focus signals Function Informing Focus status H 4 0 V L 0 5 V Connector type D SUB15 female Pin No Pin Name Pin No Pin Name 1 N C 9 N C 2 N C 10 GND 3 N C 11 N C 4...

Страница 25: ...sting camera diopter Adjust camera diopter using the following steps TIP Be sure to adjust camera diopter to prevent the following issues 1 Parfocal among objective lenses getting worse 2 Auto focusin...

Страница 26: ...set of offset values of Z stage position is recommended to be held in customer s system software If needed calibrate the set of offset values in advance with e g a mirror specimen 3 11 5 Correcting ch...

Страница 27: ...nly BXC RLI l The cable connector has a suitable orientation for the connection The connector on the wrong side cannot be connected l Use the cable provided with BXC FSU to connect BXC FSU and BXC RLI...

Страница 28: ...rity even Stop bit 1 or 2 bits switching the setting not required Terminator CR LF Flow control None The setting of the stop bits for communicating from the Host PC to BXC CBB The setting of the stop...

Страница 29: ...1 Use BXC LCBL3M or BXC LCBL6M to connect BXC RLI and BXC CBB 4 AC adapter 1 Connect the control box BXC CBB and the AC adapter provided with BXC CBB 2 Insert the power cord connector A into the conne...

Страница 30: ...3 Connect the power cord s plug C to the power outlet D on the wall 3 12 3 Connecting cables to BXC CBE1 1 Connect each user cable to BXC CBE1 as needed 26 3 12 Connecting cables 3 Setup procedures...

Страница 31: ...ctor 2 No unit connected Unit connected 3 Green Back connector 3 No unit connected Unit connected 4 2 Status display of the indicators on BXC CBE1 The following table shows the status display of each...

Страница 32: ...reference manual TIP The optimum pulse numbers for typical objectives are listed below Objective AS recommended pulse number pulses Objective AS recommended pulse number pulses Objective AS recommend...

Страница 33: ...e glare when the darkfield DF path is switched to the brightfield BF path Removing the Interlocked ND Filter TIP The ND filter has been interlocked at the factory If brightness is not enough during br...

Страница 34: ...to the original position NOTE To attach the removed ND filter again push in the frame lightly Do not push with an excessive force If the force is excessive the ND filter will be disengaged from the D...

Страница 35: ...well ventilated room Portions other than lenses Wipe the portions other than lenses with a dry soft cloth If dirt dust cannot be removed by dry wiping moisten a soft cloth with diluted neutral deterg...

Страница 36: ...becomes dark Or the brightness of the field of view is not even The light path selection lever of the trinocular tube is not stopped at the correct position Stop the light path selection lever of the...

Страница 37: ...ly engaged in the light path Turn the nosepiece until the clicking sound is heard and engage the objective in the light path The aperture diaphragm is narrowed down but the centering is not performed...

Страница 38: ...pecific hole partial segment is used Refer to To extend the life of the nosepiece and evenly apply the grease 31 Auto focusing does not work The safety lock is not attached Attach the short plug provi...

Страница 39: ...certified by one of the agencies listed in Table 1 or comprised of cordage marked with an agency marking per Table 1 or marked per Table 2 The fittings are to be marked with at least one of the agenci...

Страница 40: ...0 10 30 N V KEMA KEMA KEUR HAR 10 30 30 SEMKO AB Svenska Elektriska Materielkontrollanstalter SEMKO HAR 10 10 50 sterreichischer Verband f r Elektrotechnik VE VE HAR 30 10 50 Danmarks Elektriske Mater...

Страница 41: ...er Service Center https www olympus lifescience com support service https www olympus ims com service and support service centers Official website Official website https www olympus lifescience com ht...

Отзывы: