Olympus BX61 Скачать руководство пользователя страница 7

3

BX61

Fig. 3

1

    Getting Ready

1. A microscope is a precision instrument. Handle it with care and avoid

subjecting it to sudden or severe impact.

2. Do not use the microscope where it is subjected to direct sunlight, high

temperature and humidity, dust or vibrations. (For the operating condi-
tions, refer to chapter 5, “SPECIFICATIONS”.)

3. When moving the microscope, remove the modules that may drop

including the specimen from the microscope in advance, hold it by the
hand hook on the rear and the base as shown on the left (Fig. 3) and
carry carefully (Weight: approx. 16 kg).

# Before transporting the microscope, be sure to remove the speci-

men, eyepieces and color glass filter so that they will not drop. Also
remove other modules attached to the microscope because they
increase the system weight.

# Damage to the microscope will occur if you grasp it by the stage,

focus adjustment knob or binocular section of the observation tube.

4. The BX61 series can be used with up to two intermediate attachments

including a U-CA magnification changer, U-EPA2 eyepoint adjuster, etc.
For restrictions when using two intermediate attachments, make sure to
read the instruction manuals provided with the respective intermediate
attachments.

2

    Maintenance and Storage

1. To clean the lenses and other glass components, simply blow dirty away using a commercially available blower and

wipe gently using a piece of cleaning paper (or clean gauze).
If a lens is stained with fingerprints or oil smudges, wipe it gauze slightly moistened with commercially available absolute
alcohol.

!

Since the absolute alcohol is highly flammable, it must be handled carefully.
Be sure to keep it away from open flames or potential sources of electrical sparks –– for example, electrical
equipment that is being switched on or off, which could cause ignition of a fire.
Also remember to always use it only in a well-ventilated room.

2. Do not attempt to use organic solvents to clean the microscope components other than the glass components. To clean

them, use a lint-free, soft cloth slightly moistened with a diluted neutral detergent.

3. Do not disassemble any part of the microscope as this could result in malfunction or reduced performance.
4. When not using the microscope, set the main switch to “    ” (OFF), wait until the lamp housing has cooled down, and

cover the microscope with a dust cover before storage.

5. When disposing of the microscope, check the regulations and rules of your local government and be sure to observe

them.

3

    Caution

If the microscope is used in a manner not specified by this manual, the safety of the user may be imperiled. In addition,
the equipment may also be damaged. Always use the equipment as outlined in this instruction manual.

The warning, caution and other notes in this manual use the following symbols.

!

: Indicates that failure to follow the instructions in the warning could result in bodily harm to the

user and/or damage to equipment (including objects in the vicinity of the equipment).

# : Indicates that failure to follow the instructions could result in damage to equipment.
} : Indicates commentary (for ease of operation and maintenance).

4

    Intended use

This instrument has been designed to be used to observe magnified images of specimens in routine and research
applications. Do not use this instrument for any purpose other than its intended use.

Содержание BX61

Страница 1: ...copes Models BX61 and BX62 To ensure the safety obtain optimum performance and to familiarize yourself fully with the use of this microscope we recommend that you study this manual thoroughly before o...

Страница 2: ......

Страница 3: ...5 Selecting the Light Path of the Trinocular Tube 6 Adjusting the Tilt 1 3 Correct assembly and adjustments are indispensable to allow the microscope manifest its full performance If you are going to...

Страница 4: ...ccordance with the instruction manual may cause harmful interference to radio communications Operation of this equipment in a residential area is likely to cause harmful interference in which case the...

Страница 5: ...take the infection prevention measures The product becomes unstable if its height is increased by an accessory mounted on it In this case take anti toppling measures to prevent the specimen from being...

Страница 6: ...e touched with bare hands Before use carefully read the instruction manual Improper use could result in personal injury to the user and or damage to the equipment Indicates that the main switch is ON...

Страница 7: ...solute alcohol is highly flammable it must be handled carefully Be sure to keep it away from open flames or potential sources of electrical sparks for example electrical equipment that is being switch...

Страница 8: ...re iris diaphragm ring Page 20 Filter mount Page 9 Field iris diaphragm ring Page 19 Brightness adjustment buttons Page 8 Light path selector knob Page 17 Condenser height adjustment screw Page 19 All...

Страница 9: ...cape button Page 12 F C button Page 11 Light Preset switch Page 8 Lamp ON OFF switch Stage UP button Page 12 Stage DOWN button Page 12 Reflected lamp housing Motorized fluorescence illuminator BX RFAA...

Страница 10: ...n the light path Revolving nosepiece button Bring the specimen in focus Stage UP DOWN button focus adjustment knob Adjust the interpupillary distance Adjust the diopter Adjust the light axis Binocular...

Страница 11: ...icroscope Set the Z board or AF board For the setup and installation of the Z board refer to page 30 For the information regarding the AF board refer to the instruction manual of the AF unit Use the A...

Страница 12: ...igs 4 5 1 Press the Light Preset switch The PHOTO indicator lights up 2 Press one of the light intensity adjustment buttons to set the de sired brightness 3 Press the Light Preset switch again The ind...

Страница 13: ...the U FC filter cassette Even when a filter cassette is used a filter with thickness of no more than 3 mm can still be placed on the filter mount Using the Filter Cassette Figs 8 11 Loading Filters in...

Страница 14: ...ten the filter cassette When the filter cassette is installed the condenser may hit the filter cassette when the stage is lowered Therefore exercise caution when lowering the stage with the filter cas...

Страница 15: ...the opposite side to the stage feed knobs 1 Loosen the clamping screw and remove the focus adjustment knob 2 Remove the seal from the focus adjustment knob screw hole on the other side and attach the...

Страница 16: ...used The upper limit of stage movement has been set so that the stage stops within the coverage range of the damper spring of objective when a slide specimen thickness 1 2 mm is used When observing a...

Страница 17: ...stage by sliding each slide from the front 3 After placing the slides as far as they will go gently release the clamping lever Microscopy with Single Slide Holder Fig 18 The specimen slide can easily...

Страница 18: ...Rubber Stage Knob Caps Optional When the X Y axis knobs are fitted with the rubber caps the knobs can be adjusted without slipping and fine adjustment is possible by holding the knobs with a very lig...

Страница 19: ...lower the stage to the lower limit then remove the stage from the microscope 2 Using an Allen screwdriver loosen the substage bracket clamping screw and remove the substage 3 Press the stage UP butto...

Страница 20: ...defined double crossline can be seen in the field of view Figs 25 26 2 Looking through the right eyepiece rotate the coarse and fine focus ad justment knob to bring the specimen and double crossline...

Страница 21: ...ounterclockwise to remove it from the eyepiece and place a micrometer disk Screw the micrometer mounting frame back into the eyepiece sleeve Fig 28 Fig 29 5 Fig 29 Trinocular tube Pushed in Middle pos...

Страница 22: ...r no intermediate attachment can be combined when the BX RFAA or BX RLAA reflected light illuminator is used If you need photography using the U TBI3 also use the U TRU trinocular intermediate tube Wi...

Страница 23: ...o condenser centering screws and turn them to move the iris diaphragm image to the center of the field of view 6 Gradually open the field iris diaphragm The condenser is properly cen tered if the iris...

Страница 24: ...ure iris diaphragm ring Fig 35 2 Compatibility of Objectives and Condensers Fig 34 Fig 35 N A scale Objective Magnification Condenser Achromat aplanat U AAC Swing out U SC3 Motorized universal U UCD8A...

Страница 25: ...nly when oil is applied between the slide glass and the top surface of the condenser When oil is not present the NA is about 0 9 5 After use remove oil from the objective front lens by wiping with gau...

Страница 26: ...ht path selector knob is in an posi tioned halfway Set the knob according to the obser vation method 17 Revolving nosepiece is not correctly en gaged Make sure that the revolving nose piece clicks pro...

Страница 27: ...olving nosepiece is not corrected mounted Push slide dovetail all the way until it is stopped Objective is not correctly engaged in light path Make sure that revolving nosepiece clicks into place corr...

Страница 28: ...h sides are the same type Your view is not accustomed to micro scope observation Upon looking into eyepieces try look ing at overall field before concentrat ing on specimen range You may also find it...

Страница 29: ...1 mm fine 1 mm coarse Full stroke range 25 mm to 14 mm when the U FWT filter wheel is used 4 Revolving nosepiece 5 Observation tube Type U BI30 2 U TBI3 U TR30 2 U SWTR 3 Widefield bin ocular tube Wi...

Страница 30: ...6 5 super widefield 2X 100X for FN 26 5 super widefield 8 Operating environment Indoor use Altitude Max 2000 meters Ambient temperature 5 to 40 C 41 to 104 F Maximum relative humidity 80 for temperatu...

Страница 31: ...us m Field of view PlanN P Plan Achromat for polarized light FN22 10X 0 25 6 0 1 30 100X 28 0 2 2 20X 0 4 3 0 0 17 0 84 200X 9 3 1 1 40X 0 65 0 45 0 17 0 52 400X 2 0 0 55 100XO 1 25 0 13 0 27 1000X 0...

Страница 32: ...35 1000X 1 0 0 22 Correction collar PlanFl Plan Semi Apochromat FN26 5 20X3 340 0 75 0 55 0 17 0 45 200X 4 3 1 1 20XW3 340 0 70 0 40 0 17 0 48 200X 4 08 1 1 40X3 340 0 90 0 20 0 11 0 23 0 37 400X 2 0...

Страница 33: ...rformance we recommend that you have your Olympus representative assemble or remove this module Eyepiece WHN10X FN 22 35WHN10X FN 22 SWH10X H FN 26 5 35SWH10X FN 26 5 For the electrical connections re...

Страница 34: ...on Mounting the Z Board Fig 39 Be sure to set the main switch on the BX UCB control box to OFF before proceeding to mounting the Z board 1 Loosen the 6 clamping knobs on the 2 option slot covers on th...

Страница 35: ...ove Tighten the clamping screw while pressing downward on the cover Fig 40 Caution for Bulb Replacement During Use or Right After Use The bulb lamp socket and areas near these will be extremely hot du...

Страница 36: ...sen the condenser clamping screw using the Allen screw driver When attaching the U SC3 swing out condenser align the positioning pin at the back of the condenser with the groove in the condenser sleev...

Страница 37: ...ing Table 1 Certified Cord A power supply cord should be certified by one of the agencies listed in Table 1 or comprised of cordage marked with an agency marking per Table 1 or marked per Table 2 The...

Страница 38: ...ter sterreichischer Verband f r Elektrotechnik VE Danmarks Elektriske Materialkontroll DEMKO National Standards Authority of Ireland NSAI Norges Elektriske Materiellkontroll NEMKO Asociacion Electrote...

Страница 39: ......

Страница 40: ...ny 3500 Corporate Parkway P O Box 610 Center Valley PA 18034 0610 U S A One Corporate Drive Orangeburg NY 10962 U S A 491B River Valley Road 12 01 04 Valley Point Office Tower Singapore 248373 31 Gilb...

Отзывы: