11
BX61
3-2 Focusing Block
#If the substage is not installed in the correct position, it
may deviate from the coverage range of the damper spring
of the objective and damage the specimen.
Ensure that the top surface @ of the substage is flush with the bottom
surface of the notch ² on the mount dovetail. If they are not flush, loosen
the clamping screw ³ and lower the substage until it is stopped.
1
Replacing the Focus Adjustment Knob
(Fig. 13)
# The focus adjustment knob is installed on the right side of the micro-
scope when it left the factory.
}The focus adjustment knob is designed detachable to prevent interfer-
ence with hand during manipulation of the focus adjustment knob or X/
Y-axis knobs.
Usually attach the knob on the opposite side to the stage feed knobs.
1. Loosen the clamping screw @ and remove the focus adjustment knob
².
2. Remove the seal from the focus adjustment knob screw hole on the
other side and attach the knob by reversing the removal procedure.
3. Attach a provided seal on the screw hole | of the focus adjustment dial
³, from which the focus adjustment knob has been removed.
}The focus adjustment dial ³ can be operated with your fingertip or finger
surface at the same time as manipulating the X/Y-axis knobs.
2
F/C button
(Fig. 14)
}This button switches the movement of the stage using the focusing
knob @ or dial between F (Fine) and C (Coarse). (For the purpose of
safety, this button is set automatically to F at the moment the main
switch of the BX-UCB control box is set to “
I
” (ON).)
· Each press of the F/C button ² switches F and C alternately.
Stage Fine movement:
0.1 mm per turn
Stage Coarse movement: 1 mm per turn
Fig. 12