PHASE CONTRAST

1.

 

Remove microscope cover.

2.

 

Turn on the halogen light switch (

1

) for transmitted light to “I” (ON).

3.

 

Check the light path. The light path selector knob (

2

) should be pushed in all the way.

4.

 

Adjust the light intensity using the brightness adjustment knob (

3

). The numerals to the right

of the lamp voltage indicator LEDs (

4

) indicate the voltage.

5.

 

Adjust the interpupillar distance. While looking through the eyepieces, adjust oculars (

5

)

until the left and right fields of view coincide completely.

6.

 

Make sure the 10x phase objective (

6

) is in place (phase objective has a ‘Ph’ engraved on it).

7.

 

Place the slide with your specimen on the stage (

7

) and hold it with the specimen holder (

8

).

8.

 

Find your specimen using the stage controls (

9

).

9.

 

Focus specimen using fine/course focussing knobs (

10

).

10.

 

Adjust the diopter:

-

 

Close your left eye and focus on the specimen using the fine focus knob.

-

 

Close your right eye and focus on the specimen using the diopter ring (

11

) on the left

ocular.

-

 

Open both eyes and confirm that the focus is comfortable.

11.

 

If desired switch to the next phase objective by rotating the nosepiece (

12

) and focus.

12.

 

Focus the light on the specimen:

-

 

Close field iris diaphragm (

13

) until you can see the edges.

-

 

Focus the image of the field iris diaphragm by raising or lowering the condenser
using the condenser height adjustment knob (

14

).

-

 

Check if the circle of light is centered in the field of view. If not, use the two
condenser centering screws (

15

) to move the field iris diaphragm image to the center

of the field of view.

-

 

Open the field iris diaphragm until its image circumscribes the field of view. 

22.

 

Open the condenser aperture iris diaphragm (

16

) all the way.

23.

 

Rotate the condenser turret 

(23)

 to engage the phase contrast ring until the visible marking

matches the ‘Ph’ mark on the respective objective.

24.

 

While looking through the oculars, increase the intensity 

(3)

 until you can see clearly.

25.

 

Remove one of the oculars and look down the tube. A bright and a dark ring can be seen. If
the rings do not exactly overlap (most likely they will), use the optical element centering
screws 

(24)

 to center the phase contrast ring so that the bright ring overlaps the dark ring

within the field of view.

26.

 

Replace the ocular.

27.

 

Examine specimen and document image if necessary:

-

 

If using 35mm film, use checklist for ‘

Olympus 35mm Camera’

.

-

 

To obtain still digitized images, use checklist for ‘

Polaroid Digital Microscope

Camera (DMC).

-

 

To obtain still digitized images and digitized movies, use checklist for ‘

Sony, color

Video camera’.

This Manual: http://www.manuallib.com/file/2605418

Содержание BX51

Страница 1: ...b com ManualLib com collects and classifies the global product instrunction manuals to help users access anytime and anywhere helping users make better use of products Home http www manuallib com Chin...

Страница 2: ...3 Phase Contrast 5 Nomarski Differential Interference Contrast DIC 7 Simultaneous Fluorescence and Phase Contrast 9 Simultaneous Fluorescence and DIC 11 Polarized Light 13 APPENDIX 1 Outline Olympus B...

Страница 3: ...ve by rotating the nosepiece 12 and focus Continue until you reach the desired magnification 12 Establish Koehler illumination Close field iris diaphragm 13 until you can see the edges Focus the image...

Страница 4: ...hen finished Lower the stage by turning the focus knob 10 towards you and remove the slide with your specimen from the stage 7 Turn the nosepiece 12 back until the 10x objective is into place If the10...

Страница 5: ...ris diaphragm knob 22 the aperture iris diaphragm will get smaller 9 Examine specimen and document image if necessary If using 35mm film use checklist for Olympus 35mm Camera To obtain still digitized...

Страница 6: ...U 1 Autofluorescence observation DAPI DNA staining Hoechst 33258 33342 Chromosome WB 2 FITC Fluorescent antibody Acridine orange DNA RNA Auramine Tubercle bacillus EGFP S65T RSGFP WG 3 Rhodamine TRITC...

Страница 7: ...13 until you can see the edges Focus the image of the field iris diaphragm by raising or lowering the condenser using the condenser height adjustment knob 14 Check if the circle of light is centered i...

Страница 8: ...emove the slide with your specimen from the stage 7 Turn the nosepiece 12 back until the 10x objective for brightfield imaging is into place Turn the light intensity down using the brightness adjustme...

Страница 9: ...u can see the edges Focus the image of the field iris diaphragm by raising or lowering the condenser using the condenser height adjustment knob 14 Check if the circle of light is centered in the field...

Страница 10: ...the polarizer 25 out of the light path Pull the reflected light analyser U AN 24 out of the light path Turn the light intensity down using the brightness adjustment knob 3 Turn off the halogen light...

Страница 11: ...you the observed image and to improve the contrast pull out the aperture iris diaphragm knob 22 the aperture iris diaphragm will get smaller 9 Examine specimen and document image if necessary If usin...

Страница 12: ...ff the halogen light switch 1 to 0 OFF Cover the microscope when the mercury bulb is cooled down 15 minutes after it was turned off Leave the room CLEAN and TIDY This Manual http www manuallib com fil...

Страница 13: ...of view 8 To help adjusting the brightness of you the observed image and to improve the contrast pull out the aperture iris diaphragm knob 22 the aperture iris diaphragm will get smaller 9 Examine sp...

Страница 14: ...path Pull the polarizer 25 out of the light path Pull the reflected light analyser U AN 24 out of the light path Turn the light intensity down using the brightness adjustment knob 3 Turn off the halog...

Страница 15: ...light on the specimen Close field iris diaphragm 13 until you can see the edges Focus the image of the field iris diaphragm by raising or lowering the condenser using the condenser height adjustment k...

Страница 16: ...x objective was used clean the lens carefully with lens paper ONLY do not use Kimwhipes Pull the polarizer 25 out of the light path Pull the reflected light analyser U AN 24 out of the light path Turn...

Страница 17: ...15 APPENDIX 1 Outline Olympus BX51 Microscope This Manual http www manuallib com file 2605418...

Отзывы: