SIMULTANEOUS FLUORESCENCE AND PHASE CONTRAST

1.

 

Remove the microscope cover.

2.

 

To turn on the mercury bulb, set the main switch 

(18)

 on the Power Supply Unit to “I”(ON).

The light indicator 

(19) 

on the Power Supply Unit should go ON. Wait 5 to 10 minutes for

the mercury bulb to stabilize. The mercury bulb should stay on for at least 15 minutes before
turning off.

3.

 

Setup the microscope for Phase Contrast imaging (see instructions on ‘

Phase contrast

 

page

5, #2-26).

4.

 

Engage the fluorescence mirror unit 

(17) 

until the proper filter cube is in place:

-

 

WU – #1, dichroic mirror DM400, excitation filter BP330-385, barrier filter BA420

-

 

WB – #2, dichroic mirror DM500, excitation filter BP450-480, barrier filter BA515

-

 

WG – #3, dichroic mirror DM570, excitation filter BP510-550, barrier filter BA590

5.

 

To allow light from the mercury bulb to reach the specimen, slide the shutter knob 

(20)

 to

position marked 

.

6.

 

Adjust the transmitted light for the best balance of fluorescence and phase contrast brightness
using the brightness adjustment knob 

(3)

.

7.

 

To improve image contrast and to prevent color fading of fluorescent light in other part than
observed region, pull out the field iris diaphragm knob 

(21)

 so that the image of the field iris

diaphragm just circumscribes the field of view.

8.

 

To help adjusting the brightness of you the observed image and to improve the contrast, pull
out the aperture iris diaphragm knob 

(22)

, the aperture iris diaphragm will get smaller.

9.

 

Examine specimen and document image if necessary:

-

 

If using 35mm film, use checklist for ‘

Olympus 35mm Camera’

.

-

 

To obtain still digitized images, use checklist for ‘

Polaroid Digital Microscope

Camera (DMC).

-

 

To obtain still digitized images and digitized movies, use checklist for ‘

Sony, color

Video camera’.

10.

 

To avoid bleaching, slide the shutter knob 

(20)

 to the position marked 

˜

 whenever you are

not observing the specimen.

11.

 

When finished:

-

 

Slide the shutter knob

 (20)

 to the position marked 

˜

.

-

 

If mercury bulb has burned for at least 15 minutes, turn off the mercury bulb by
setting the main switch 

(18)

 on the Power Supply Unit to “O”(OFF). The light

indicator 

(19) 

on the Power Supply Unit should go out.

NOTE:

 The mercury bulb needs to cool down for at least 15 minutes before turning

back on.

-

 

Lower the stage by turning the focus knob 

(10)

 towards you and remove the slide

with your specimen from the stage 

(7)

.

-

 

Turn the nosepiece (

12

) back until the 10x objective for brightfield imaging is into

place.

-

 

Push the field iris diaphragm knob 

(21)

 and the aperture iris diaphragm knob 

(22)

 al

the way back into the microscope.

-

 

Engage the fluorescence mirror unit 

(17) 

back to BF (brightfield) - #6

-

 

Rotate the condenser turret 

(23)

 until the BF (brightfield) mark is visible.

-

 

Turn the light intensity down using the brightness adjustment knob (

3

).

This Manual: http://www.manuallib.com/file/2605418

Содержание BX51

Страница 1: ...b com ManualLib com collects and classifies the global product instrunction manuals to help users access anytime and anywhere helping users make better use of products Home http www manuallib com Chin...

Страница 2: ...3 Phase Contrast 5 Nomarski Differential Interference Contrast DIC 7 Simultaneous Fluorescence and Phase Contrast 9 Simultaneous Fluorescence and DIC 11 Polarized Light 13 APPENDIX 1 Outline Olympus B...

Страница 3: ...ve by rotating the nosepiece 12 and focus Continue until you reach the desired magnification 12 Establish Koehler illumination Close field iris diaphragm 13 until you can see the edges Focus the image...

Страница 4: ...hen finished Lower the stage by turning the focus knob 10 towards you and remove the slide with your specimen from the stage 7 Turn the nosepiece 12 back until the 10x objective is into place If the10...

Страница 5: ...ris diaphragm knob 22 the aperture iris diaphragm will get smaller 9 Examine specimen and document image if necessary If using 35mm film use checklist for Olympus 35mm Camera To obtain still digitized...

Страница 6: ...U 1 Autofluorescence observation DAPI DNA staining Hoechst 33258 33342 Chromosome WB 2 FITC Fluorescent antibody Acridine orange DNA RNA Auramine Tubercle bacillus EGFP S65T RSGFP WG 3 Rhodamine TRITC...

Страница 7: ...13 until you can see the edges Focus the image of the field iris diaphragm by raising or lowering the condenser using the condenser height adjustment knob 14 Check if the circle of light is centered i...

Страница 8: ...emove the slide with your specimen from the stage 7 Turn the nosepiece 12 back until the 10x objective for brightfield imaging is into place Turn the light intensity down using the brightness adjustme...

Страница 9: ...u can see the edges Focus the image of the field iris diaphragm by raising or lowering the condenser using the condenser height adjustment knob 14 Check if the circle of light is centered in the field...

Страница 10: ...the polarizer 25 out of the light path Pull the reflected light analyser U AN 24 out of the light path Turn the light intensity down using the brightness adjustment knob 3 Turn off the halogen light...

Страница 11: ...you the observed image and to improve the contrast pull out the aperture iris diaphragm knob 22 the aperture iris diaphragm will get smaller 9 Examine specimen and document image if necessary If usin...

Страница 12: ...ff the halogen light switch 1 to 0 OFF Cover the microscope when the mercury bulb is cooled down 15 minutes after it was turned off Leave the room CLEAN and TIDY This Manual http www manuallib com fil...

Страница 13: ...of view 8 To help adjusting the brightness of you the observed image and to improve the contrast pull out the aperture iris diaphragm knob 22 the aperture iris diaphragm will get smaller 9 Examine sp...

Страница 14: ...path Pull the polarizer 25 out of the light path Pull the reflected light analyser U AN 24 out of the light path Turn the light intensity down using the brightness adjustment knob 3 Turn off the halog...

Страница 15: ...light on the specimen Close field iris diaphragm 13 until you can see the edges Focus the image of the field iris diaphragm by raising or lowering the condenser using the condenser height adjustment k...

Страница 16: ...x objective was used clean the lens carefully with lens paper ONLY do not use Kimwhipes Pull the polarizer 25 out of the light path Pull the reflected light analyser U AN 24 out of the light path Turn...

Страница 17: ...15 APPENDIX 1 Outline Olympus BX51 Microscope This Manual http www manuallib com file 2605418...

Отзывы: