![Olympus BX-RLAA Скачать руководство пользователя страница 13](http://html1.mh-extra.com/html/olympus/bx-rlaa/bx-rlaa_instructions-manual_3904847013.webp)
BX-RLAA
10
4
Using the Interlocked ND Filter
(Fig. 5)
}The ND filter is interlocked with the brightfield (BF) light path switching and
makes it possible to reduce glare when the darkfield (DF) path is switched
to the brightfield (BF) path.
Removing the Interlocked ND Filter
}The ND filter has been interlocked at the factory. If brightness is not enough
during brightfield, DIC or other observation, remove the ND filter as de-
scribed below.
1. Select the DF observation path, and remove the interlocked ND filter cover
by loosening the retaining screw @ screw using the Allen screwdriver.
2. Remove the ND filter frame ² which is visible after the cover removal.
3. Attach the cover to the original position.
# To attach the removed ND filter again, push in the frame lightly. Do not
push with an excessive force. If the force is excessive, the ND filter will
be disengaged fro the DF path and the main switch of the BX-UCB
should be set to “
I
” (ON) then “ ” (OFF) again.
Fig. 5
@
²