background image

15

SIMULTANEOUS FLUORESCENCE OBSERVATION

METHODS

}By properly combining equipment, this system can be used in transmitted light brightfield observation, transmitted phase

contrast observation and transmitted light DIC observation in addition to the reflected fluorescence observation. With speci-
mens  that  fade  rapidly,  fading  can  be  minimized  by  initially  using  transmitted  light  phase  contrast  or  transmitted  DIC
observation for positioning. Reflected fluorescence observation can also be executed simultaneously with phase contrast
or DIC observation, making it easy to tell which portion of the specimen is fluorescing.

1

    Simultaneous Reflected Fluorescence and Phase Contrast Observations

The phase contrast observation requires a phase contrast condenser (U-PCD2) or a universal condenser (U-UCD8 or U-
UCD8A) and a Ph objective.

1. Engage a dummy mirror unit (an empty position on the turret) in the light path.
2. Rotate the phase contrast turret to show the same number as the Ph number shown on the objective.
3. Adjust the optical axis between the ring slit and phase plate by centering them.
4. Engage the mirror unit corresponding to the desired excitation into the light path and open the shutter.
5. Adjust the transmitted light for the best balance of fluorescence and phase contrast brightness.
}Use ND filters or the intensity control button on the microscope base to adjust the transmitted light intensity.
}For details on using phase contrast observation, refer to the instruction manual provided with the phase contrast con-

denser or universal condenser.

2

    Simultaneous Reflected Fluorescence and Transmitted Nomarski DIC Observations

The transmitted light Nomarski DIC (Differential Interference Contrast) observation requires the following accessories:

1) Universal condenser (U-UCD8 or U-UCD8A)
2) Transmitted DIC slider (U-DICT, U-DICTS, U-DICTHR or U-DICTHC)
3) Analyzer (U-AN or U-AN360-3)
4) Revolving nosepiece or motorized revolving nosepiece for DIC

}In order for reflected fluorescence to be effective in the simultaneous observation, insert the analyzer (U-AN or U-AN360-

3) into the analyzer insertion slot above the dichroic mirror on the illuminator.
Do not insert the U-ANT analyzer in the transmitted light DIC slider, for this will dim the fluorescence observation image
and cause the analyzer to be burnt.

1. Engage the dummy mirror unit (or an empty position) in the light path.
2. Adjust the polarizer on the universal condenser to the “crossed Nicol” (complete extinction) status.
3. Insert the transmitted light DIC slider into the position provided on the nosepiece.
4. Rotate the turret on the universal condenser to select the Nomarski prism matching the objective to be used for observa-

tion.

5. Engage the objective to be used in the light path.
6. Place the specimen on the stage and focus on the specimen.
7. Adjust the field iris diaphragm of the transmission light illuminator (built in the microscope base) and the aperture iris

diaphragm of the universal condenser.

8. Turn the prism movement knob on the transmitted light DIC slider to adjust contrast of the DIC image.
9. Engage the mirror unit corresponding to the desired excitation and open the shutter.

Adjust the transmitted light for optimum fluorescence and DIC image brightness.

}For details on the transmitted DIC observation, refer to the instruction manual provided with the U-UCD8 or U-UCD8A

universal condenser.

10.

Reference

}We recommend the use of the highly wear-resistant U-ANH analyzer slider instead of the U-AN analyzer when you are

frequently switching between reflected fluorescence observation and transmitted light Nomarski DIC observation and
need to use both observations simultaneously.

}However, if you are frequently switching between reflected fluorescence observation and transmitted light Nomarski DIC

observation but you do not need to use both simultaneously, then it will be more convenient for you to use the M-DICT3
DIC mirror unit instead of an analyzer (U-AN or U-ANH). This facilitates the switching operation because the analyzer
simultaneously enters the light path when the fluorescence mirror unit is switched to the DIC mirror unit.

Содержание BX-RFAA

Страница 1: ...urself fully with the use of this system we recommend that you study this manual thoroughly before operating the system Retain this instruction manual in an easily accessible place near the work desk...

Страница 2: ...UIDE 6 SPECTRAL CHARACTERISTICS OF FILTERS 7 SPECIFICATIONS 4 5 6 7 8 14 15 16 17 18 21 22 1 General Precautions for Observation 8 2 Selecting of the Fluorescence Mirror Unit 8 9 3 Objectives for Vari...

Страница 3: ...35 PROPER SELECTION OF THE POWER SUPPLY CORD 36 37 8 OPTIONAL MODULES 23 28 1 6 Position Filter Slider U RSL6 23 24 2 6 Position Barrier Filter Slider U RSL6EM 25 3 Rectangular Field Stop Unit BX RFS...

Страница 4: ...ed please select the proper power cord by referring to the chapter PROPER SELECTION OF THE POWER SUPPLY CORD at the end of this instruction manual If the proper power cord is not used product safety a...

Страница 5: ...ily accessed and unplugged in case of emergency 2 Maintenance and Storage 1 Clean all glass components by wiping gently with gauze To remove fingerprints or oil smudges wipe with gauze slightly moiste...

Страница 6: ...tion manual The following symbols are used to set off text in this instruction manual Indicates that failure to follow the instructions in the warning could result in bodily harm to the user and or da...

Страница 7: ...plate Mirror unit indication pockets Page 31 x 6 Field iris diaphragm knob Page 11 Aperture iris diaphragm knob Page 12 Collector lens focusing knob Page 13 Burner centering knobs Page 13 Mirror focus...

Страница 8: ...ate an original fluorescence mirror unit it is recom mended to use the U MF2 empty mirror unit which does not include filter page 32 A blank indication stickers are provided with the illuminator Use t...

Страница 9: ...stabilizes 5 to 10 minutes Main switch P 10 Place the specimen on the stage Specimen holder X Y axis knobs Engage the mirror unit matching the speci men MU Mirror Unit Right or MU Mirror Unit Left bu...

Страница 10: ...st it near your microscope Mercury Lamp Housing U LH100HGAPO U LH100HG Motorized Reflected Fluorescence Illuminator BX RFAA Motorized Revolving Nosepiece Microscope BX61 BX62 etc Control Box BX UCB Ha...

Страница 11: ...deteriorate In such a case slightly reduce the excitation light intensity to slow color fading down and improve the fluorescence image To reduce the excitation light intensity use ND filters or apertu...

Страница 12: ...ne orange staining DNA RNA Auramine staining Tubercle bacteria EGFP S65T RSGFP IB U MWIB2 U MNIB2 DM505 BP460 490 BP470 490 BA510IF U MWG2 BP510 550 G U MNG2 DM570 BP530 550 BA590 U MSWG2 BP480 550 Rh...

Страница 13: ...0 minutes after ignition The mercury burner may not be ignited from the beginning due to variance between individual products In this case set the main switch to OFF wait for 5 to 10 seconds then set...

Страница 14: ...in the two field iris diaphragm centering screws and adjust them so that the field iris image comes at the center of the field of view 7 While pushing in the field iris diaphragm knob enlarge the fie...

Страница 15: ...Pull out the aperture iris diaphragm knob to minimize the aperture iris diameter 8 Pull out the field iris diaphragm knob to minimize the field iris dia phragm Now the aperture iris image should be vi...

Страница 16: ...to the center of field of view 4 Engage the empty place the objective cap should be removed in the light path 5 Pull out the aperture iris diaphragm knob to minimize it and push in the field iris diap...

Страница 17: ...centering of the mirror arc image 8 Using the ND Filters Fig 5 Specimen color fading can be delayed by reducing the excitation light intensity with ND filters Use the ND filters as far as they do not...

Страница 18: ...r reflected fluorescence to be effective in the simultaneous observation insert the analyzer U AN or U AN360 3 into the analyzer insertion slot above the dichroic mirror on the illuminator Do not inse...

Страница 19: ...ey circumscribe field of view Fluorescence mirror unit in use does not match specimen Use fluorescence mirror unit match ing specimen c Field of view is obscured or not evenly illuminated Objective is...

Страница 20: ...cted Connect firmly e Hand switch or PC indicates that shutter is released but light is not incident or image is cut off Shutter is moved due to external vibra tions Try perform shutter operation agai...

Страница 21: ...BV excitation Wide band 5 BV excitation Narrow band 6 B excitation Wide band U MWU2 U MNU2 U MNV2 U MWBV2 U MNBV2 U MWB2 Transmittance Transmittance Transmittance Wavelength nm Wavelength nm Waveleng...

Страница 22: ...B excitation Wide band 10 IB excitation Narrow band 11 G excitation Wide band 12 G excitation Narrow band U MNB2 U MSWB2 U MWIB2 U MNIB2 U MWG2 U MNG2 Transmittance Transmittance Wavelength nm Wavelen...

Страница 23: ...and 16 U excitation for color separation Narrow band 17 IB excitation for color separation Wide band 18 IB excitation for color separation Narrow band U MSWG2 U MWIG2 U MWIY2 U MNUA2 U MWIBA2 U MNIBA2...

Страница 24: ...m having a specific wavelength is selected from a wide spectrum of wavelengths The five major peaks of luminance are at wavelengths of 365 366 404 7 435 546 1 and 577 0 579 1 nm In addition light beam...

Страница 25: ...Available observations Reflected fluorescence Reflected fluorescence Transmitted light DIC Reflected fluorescence Phase contrast Transmitted light Applicable microscope frames BX41 BX51 BX52 BX61 BX6...

Страница 26: ...mounted filters include an exciter filter insert it so that the arrow inscription on the side faces down a a Filter with thickness of 4 mm or more Place each filter so that it fits in side the holder...

Страница 27: ...e rear panel is used avoid using the interference type or color glass type filters This is because the 6 position filter slider near the rear panel is one of the positions where the energy from the li...

Страница 28: ...ially available round stickers in the same way as with the U RLS6 filter slider 4 Gently insert the filter slider from the right of the analyzer insertion slot on the illuminator and attach the slider...

Страница 29: ...op into the position of the field iris diaphragm then tighten the clamping screw Operation Procedure 1 Insert the provided adjustment knobs into the two adjustment screw holes near the front panel and...

Страница 30: ...he Allen screwdriver Operation Procedure Observing a Double Stained Specimen 1 Set up normal reflected fluorescence observation 2 Mount the fluorescence mirror units for double staining and engage the...

Страница 31: ...luorescence of shorter wavelengths near blue Observing a Triple Stained Specimen The operation is basically similar to the double stained specimens but fluorescence mirror units for triple staining sh...

Страница 32: ...ator is used only for clamping the screws inside the illuminator To retain the performance have your local Olympus representative conduct this work NOTES Be sure to insert the sliders in the orientati...

Страница 33: ...volving nosepiece as far as it will go to connect it to the motorized revolving nosepiece connector U D6REO P5REMC Remove the screw which has been clamped at the factory using a Phillips screwdriver T...

Страница 34: ...ighten the clamping screw firmly If the clamping screw is loose the turret will be unable to be rotated due to interference with the turret cover 4 Check the mount dovetail number and place upside dow...

Страница 35: ...unit frame Dimensions of Optical Parts Barrier filter Excitation filter Dichroic mirror 26 0 1 0 3 x 38 0 1 0 3 mm thickness 1 0 05 mm Diameter 25 0 1 0 2 mm max thickness 6 mm Barrier filter commerc...

Страница 36: ...the two burner holding screws 4 Attach the positive pole of a specified mercury burner to the fixed mount on the upper side then the negative pole to the mount on the lower side Be sure to use the US...

Страница 37: ...air the safety f the equipment replace the burner when it has been used for 200 USH102D or 300 HBO103W 2 hours The burner may crack if used beyond the specified life time 2 Before replacing the burner...

Страница 38: ...he input power supply match the requirements inscribed on the rating plate The 100 V system can be used with voltages in the 100 to 120 V range and 200 V system can be used with voltages in the 220 to...

Страница 39: ...e coupling Table 1 Certified Cord A power supply cord should be certified by one of the agencies listed in Table 1 or comprised of cordage marked with an agency marking per Table 1 or marked per Table...

Страница 40: ...thority of Ireland NSAI Norges Elektriske Materiellkontroll NEMKO Asociacion Electrotecnica Y Electronica Espanola AEE Hellenic Organization for Standardization ELOT Instituto Portugues da Qualidade I...

Страница 41: ...MEMO...

Страница 42: ...buya ku Tokyo Japan Postfach 10 49 08 20034 Hamburg Germany 2 Corporate Center Drive Melville NY 11747 3157 U S A 491B River Valley Road 12 01 04 Valley Point Office Tower Singapore 248373 2 8 Hondura...

Отзывы: