DMTA-10045-01EN, Rev. E, August 2016
Applications
171
2.
Lift the probe into the air, then press the MAIN menu key (
) and adjust the
ANGLE
(E key) so that the dot moves upwards in the XY view (see Figure 6-66 on
Figure 6
‑
66 Adjusting the angle to move the dot upwards
3.
If necessary, press the GAIN key (
) and adjust the
GAIN
to keep the air
signal dot on the screen (see Figure 6-67 on page 172).
The MIA probes exhibit a very high sensitivity from 10 kHz to 18 kHz, and therefore
you should expect to use much lower gains; for example, around 25 dB.
NOTE
Содержание BondMaster 600
Страница 8: ...DMTA 10045 01EN Rev E August 2016 Table of Contents viii ...
Страница 10: ...DMTA 10045 01EN Rev E August 2016 List of Abbreviations x ...
Страница 26: ...DMTA 10045 01EN Rev E August 2016 Important Information Please Read Before Use 16 ...
Страница 28: ...DMTA 10045 01EN Rev E August 2016 Introduction 18 ...
Страница 32: ...DMTA 10045 01EN Rev E August 2016 Chapter 1 22 ...
Страница 72: ...DMTA 10045 01EN Rev E August 2016 Chapter 3 62 ...
Страница 78: ...DMTA 10045 01EN Rev E August 2016 Chapter 4 68 Figure 4 2 Crosshairs and the NULL point ...
Страница 130: ...DMTA 10045 01EN Rev E August 2016 Chapter 5 120 ...
Страница 141: ...DMTA 10045 01EN Rev E August 2016 Applications 131 Figure 6 13 RUN 3 XY FLY DOT Figure 6 14 RUN 4 XY SCAN ...
Страница 216: ...DMTA 10045 01EN Rev E August 2016 Chapter 7 206 ...
Страница 230: ...DMTA 10045 01EN Rev E August 2016 Appendix A 220 ...
Страница 234: ...DMTA 10045 01EN Rev E August 2016 Appendix B 224 ...
Страница 240: ...DMTA 10045 01EN Rev E August 2016 List of Figures 230 ...
Страница 242: ...DMTA 10045 01EN Rev E August 2016 List of Tables 232 ...