4.3 Using EndoTherapy accessories
71
BF-190 Series OPERATION MANUAL
Ch.4
Bronchoalveolar lavage
Using the BAL (bronchoalveolar lavage) kit
Using a syringe
CAUTION
Angled or incomplete insertion may result in leakage of solution from the valve.
1
Disconnect the suction tube from the suction
valve. Connect the suction tube to the suction
connector of a commercially available BAL kit.
Connect the BAL kit’s suction line to the suction
connector of the suction valve.
Figure 4.10
2
Securely insert a syringe filled with lavage fluid (e.g., saline) into the slit of the biopsy
valve and press the plunger to feed lavage fluid.
3
Depress the suction valve to aspirate lavage fluid.
1
Securely insert a syringe into the slit of the biopsy valve.
2
Press the plunger to feed lavage fluid.
Figure 4.11
3
With the syringe attached, slowly withdraw the plunger to aspirate lavage fluid.
Suction tube
BAL kit
Feed fluid
Aspirate