3.6 Inspection of ancillary equipment
BF-190 Series OPERATION MANUAL
Ch.3
Connect the ancillary equipment to the endoscope as described below.
Connection to the light source
WARNING
• If the endoscope connector and light source are not connected properly, the
endoscopic image may flicker or may not be displayed. Continuous use of such an
endoscope may cause patient injury, bleeding, and/or perforation.
• Firmly connect the suction tube from the suction pump to the suction connector on
the suction valve. If the suction tube is not attached properly, debris may drip from
the tube. The patient, operator and/or equipment could be contaminated and
equipment malfunction can result.
CAUTION
Before connecting the endoscope connector to the light source, confirm that the
endoscope connector, including the electrical contacts, is completely dry and
foreign objects such as detergent remnants, hard water residue, finger grease,
dust, and lint are not on the electrical contacts. If the endoscope is used with the
electrical contacts wet and/or dirty, the endoscope and/or light source may
malfunction.
3.6
Inspection of ancillary equipment
Inspect the following equipment as described in their respective instruction manuals.
• Light source
• Video system center
• Monitor
• Suction pump
• EndoTherapy accessories
3.7
Connection of the endoscope and ancillary
equipment
Содержание BF-1TH190
Страница 2: ......
Страница 6: ...iv Contents BF 190 Series OPERATION MANUAL...
Страница 22: ...16 1 1 Checking the package contents BF 190 Series OPERATION MANUAL Ch 1...
Страница 32: ...26 2 2 Specifications BF 190 Series OPERATION MANUAL Ch 2...
Страница 56: ...50 3 8 Inspection of the endoscopic system BF 190 Series OPERATION MANUAL Ch 3...
Страница 80: ...74 4 5 Transportation of the endoscope BF 190 Series OPERATION MANUAL Ch 4...
Страница 88: ...82 5 4 Returning the endoscope for repair BF 190 Series OPERATION MANUAL Ch 5...