DMTA-10022-01EN, Rev. C, January 2015
Using Special Functions
135
2.
In the
ALARM
screen (see Figure 7-7 on page 135):
a)
Set
ALARM ENABLE
to
ON
to activate the alarm function.
b)
In
ALARM MODE
, select the desired alarm mode (
STANDARD
,
B-SCAN
,
or
REDUCTION RT
[reduction rate]).
The other parameters vary depending on the alarm mode selection.
The
B-SCAN
option only appears when the B-scan mode is active (see “Using the
REDUCTION RT
option only appears when the active file is configured with the
FILE DATA MODE
parameter set to
REDUCTION RT
Figure 7-7 Setting up the STANDARD alarm
3.
When
ALARM MODE
is set to
STANDARD
or
B-SCAN
, set the
LOW ALARM
and the
HIGH ALARM
values.
OR
When
ALARM MODE
is set to
REDUCTION RT
, set the
YELLOW ALARM
and
the
RED ALARM
values.
4.
Press
[MEAS]
to return to the measurement screen.
NOTE
Содержание 45MG
Страница 8: ...DMTA 10022 01EN Rev C January 2015 Table of Contents viii List of Tables 227 Index 229...
Страница 26: ...DMTA 10022 01EN Rev C January 2015 Introduction 16...
Страница 36: ...DMTA 10022 01EN Rev C January 2015 Chapter 1 26...
Страница 42: ...DMTA 10022 01EN Rev C January 2015 Chapter 2 32...
Страница 136: ...DMTA 10022 01EN Rev C January 2015 Chapter 6 126...
Страница 150: ...DMTA 10022 01EN Rev C January 2015 Chapter 7 140...
Страница 157: ...DMTA 10022 01EN Rev C January 2015 Configuring the Instrument 147 Figure 8 3 The COMM screen...
Страница 158: ...DMTA 10022 01EN Rev C January 2015 Chapter 8 148...
Страница 196: ...DMTA 10022 01EN Rev C January 2015 Chapter 10 186...
Страница 208: ...DMTA 10022 01EN Rev C January 2015 Chapter 11 198...
Страница 226: ...DMTA 10022 01EN Rev C January 2015 Appendix A 216...
Страница 232: ...DMTA 10022 01EN Rev C January 2015 Appendix C 222...