DMTA-10004-01EN, Rev. D, November 2016
Basic Operation
77
current ID. The ID is automatically incremented for the next measurement. When you
press
[FILE]
, the name of the active file appears in the ID bar above the menu (see
Figure 5
‑
10 The active file name appearing in the ID bar
The NONAME00 increment type file, starting with the 001 ID, is the active file by
default when you first use the 38DL PLUS or after resetting the instrument memory.
You can create various types of files and define IDs to represent various 1-D, 2-D, or
3-D thickness measurement locations. When restarting, the instrument automatically
opens the last file that was used.
The following special cases may occur:
•
When the thickness value is blank, “——-” is saved rather than a value.
•
When a measurement is already stored at the current ID, the new value
overwrites the old thickness reading unless the overwrite protection is enabled
(see “Setting the ID Overwrite Protection” on page 196).
•
When the ID increment reaches the end of a sequence and cannot be updated,
Last ID
appears on the help text bar, a long beep sound is emitted (when the
beeper is active), and the ID on the display remains unchanged.
Refer to “Using the Datalogger” on page 159 for more information on the datalogger.
To save data at the current ID in the active file
While the desired thickness value and waveform are displayed, press
[SAVE/SEND]
to save the measured thickness value.
OR
Name of the active file
Содержание 38DL PLUS
Страница 9: ...DMTA 10004 01EN Rev D November 2016 Table of Contents ix Index 297 ...
Страница 10: ...DMTA 10004 01EN Rev D November 2016 Table of Contents x ...
Страница 28: ...DMTA 10004 01EN Rev D November 2016 Important Information Please Read Before Use 16 ...
Страница 30: ...DMTA 10004 01EN Rev D November 2016 Introduction 18 ...
Страница 50: ...DMTA 10004 01EN Rev D November 2016 Chapter 2 38 ...
Страница 60: ...DMTA 10004 01EN Rev D November 2016 Chapter 3 48 ...
Страница 102: ...DMTA 10004 01EN Rev D November 2016 Chapter 6 90 ...
Страница 221: ...DMTA 10004 01EN Rev D November 2016 Using the Datalogger 209 f Press MEAS to exit the minimum review ...
Страница 222: ...DMTA 10004 01EN Rev D November 2016 Chapter 11 210 ...
Страница 254: ...DMTA 10004 01EN Rev D November 2016 Chapter 13 242 ...
Страница 274: ...DMTA 10004 01EN Rev D November 2016 Chapter 14 262 ...
Страница 286: ...DMTA 10004 01EN Rev D November 2016 Chapter 15 274 ...
Страница 294: ...DMTA 10004 01EN Rev D November 2016 Appendix A 282 ...
Страница 300: ...DMTA 10004 01EN Rev D November 2016 Appendix C 288 ...
Страница 306: ...DMTA 10004 01EN Rev D November 2016 List of Figures 294 ...