stand for 4 minutes. The protection mode can be activated automatically from 9 successive starts.
The curing light uses a beep to indicate the following situations:
1 -
Activation of Selection button;
2 -
Activation of the ON/OFF button;
3 -
After operation for 10 seconds;
4 -
After operation for 20 seconds.
Before starting the curing light use, install the optical fiber tip (1) and the light hood (2).
To install the tip: first, attach the light hood (2) to the tip and then attach it to the curing light unit (3). The
tip must be pushed all the way to the fitting end.
To activate/deactivate the curing light: Press the On/Off switch (5) twice. The activation LED (4) will light
up. The curing light unit will run for 20 seconds and will turn Off automatically.
To toggle the operating mode: Each tap on the selection button (9) switches the operation mode. Press the
selection button (9) until the orange LED indicates the desired mode.
The curing light operates in the following modes:
- Continuous (6): LED on the maximum power for 20 seconds;
- Ramp (7): LED activates at its minimum power, gradually increasing. In 5 seconds reaches its maximum
power until the cycle of 20 seconds is finished;
- Pulse (8): LED keeps blinking for 20 seconds.
5.15.1
- Precautions on the Curing Light Use
- Prohibited the use in patients who present biological reactions of sensitivity to light;
- Do not aim the light of the curing light directly at the eyes. The curing light produces optical radiation
emitted by LED;
- Do not touch the tip directly on the curing material. This will prevent material adhering to the tip,
impairing the device performance;
- Do not use the curing light without the light hood;
- The optical fiber tip should only be used on the teeth. Avoid touching the patient's gums, lips or skin;
- Use the curing light only on the teeth;
- After 40 seconds of continuous use, the tip end can reach 56º C;
- Consecutive activations of the curing light with optical fiber tip may cause the tip end to heat, reaching
a maximum temperature of 68º C.
- Brushless Electric Micromotor - Side Panel (Optional)
Technical Features:
- Power Supply: 24 VCA;
- Power Rating: ~33 W.
- Rotation Speed:
a)
20:1 from 100 to 2.000 rpm;
b)
1:1 from 2.000 rpm to 40.000 rpm;
c)
1:5 from 10.000 to 200.000 rpm;
- Torque: ~3,5 Ncm;
- LED: 25.000 Lux.
Before starting the operation, connect the
handpiece to the electrical micromotor (15).
To activate the rotation: press the
propulsion/activation pedal (section 5.5) or the
Manual Start key (11). The On/Off indicator (2) will
turn blue.