Olsen Vet Cart Premium Скачать руководство пользователя страница 1

 

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

User’s Manual

 

Vet Cart Premium 

 

Содержание Vet Cart Premium

Страница 1: ...User s Manual Vet Cart Premium...

Страница 2: ......

Страница 3: ...Tank Drainage 7 5 3 3 Pressure Regulator for Instruments 7 5 6 1 Handpieces Arrangement on the Working Table 8 5 6 2 Instruments Control Panel 9 5 8 1 Ultrasonic Scaler Tips 10 5 8 2 Ultrasonic Scaler...

Страница 4: ...nfection 18 10 Technical Features 21 11 Symbology 23 12 Dimensional 24 13 Important Notes 25 13 4 1 Contraindications on the Ultrasonic Scaler Use 27 13 4 2 Curing Light Contraindications 27 14 Troubl...

Страница 5: ...in the case of the customer order 2 Handpieces Dental instruments acquired with the equipment described in the purchase order and invoice of the product according to customer request 3 Equipment Prese...

Страница 6: ...have very low noise level in the displacement Before Turning On the Equipment 1 Check if the equipment is properly installed in accordance to the instructions in the Chapter 7 Installation Requiremen...

Страница 7: ...the airflow in the automatic drain adjustment valve can impair the performance of pneumatic and saliva ejector instruments 5 3 2 Air Tank Drainage The air tank drain is a preventive procedure which a...

Страница 8: ...3 X ray Viewer 4 Instruments Holder 5 Working Table Handle The working table s function is to provide handpieces for performing veterinary procedures Note Do not use high speed instruments on low spee...

Страница 9: ...use of specific tools The optical fiber handpiece coupling 4 is optionally supplied with special Midwest coupling with 3 1 V output for handpieces with optical fiber lighting Note The pneumatic system...

Страница 10: ...reading it carefully then insert the Torque wrench 4 into the tip and then rotate it clockwise until it is tight To remove the tip turn it counterclockwise with the Torque wrench After installing the...

Страница 11: ...ients with cardiac pacemakers or anesthetized During the ultrasonic scaler use the tip temperature may rise if the spray is not used Always use enough spray water for the cooling of the handpiece and...

Страница 12: ...ctor 2 from the holder 5 To deactivate the suction Place the ejector 2 in its holder 5 X Ray Viewer The x ray viewer is an LED illumination display that allows you to view x rays without the need to m...

Страница 13: ...the Prophy Jet Clean the sodium bicarbonate tank at the end of each procedure 5 14 3 Bicarbonate Jet and Prophy Jet Use Precautions Do not fill bicarbonate reservoirs with more than the indicated amo...

Страница 14: ...le of 20 seconds is finished Pulse 8 LED keeps blinking for 20 seconds 5 15 1 Precautions on the Curing Light Use Prohibited the use in patients who present biological reactions of sensitivity to ligh...

Страница 15: ...ptions use the adjustment keys 13 to set the parameter To adjust pneumatic activation mode use the adjustment keys 13 to set the parameter The selected pneumatic activation method is shown on the disp...

Страница 16: ...inting It is applied to all structural metal parts of the equipment The special polyurethane paint provides the equipment with high durability coating and has antibacterial properties according to JIS...

Страница 17: ...isk of damage to the equipment The contact with clean and filtered water offers no known risk to the user or the patient but the use of sanitized gloves is recommended to fill the tanks and avoid cros...

Страница 18: ...work for installation and maintenance or contact us by e mail export3 olsen odo br or if you prefer 55 48 2106 6000 Installation should be made only by an authorized technician Installation performed...

Страница 19: ...carded Curing Light To clean and disinfect the curing light tips use gauze or disposable wipe moistened with soap and neutral detergent Do not use alcohol strong or abrasive alkaline detergents deterg...

Страница 20: ...ds collector cover 1 2 Remove the solids collector 2 for cleaning 3 After cleaning re assemble the solids collector cover 1 Daily Procedure at Working Day At the end of the working day observe the fol...

Страница 21: ...Continuous operation Thermo Comfort system ultrasonic scaler optical fiber coupling lighting x ray viewer and electrical micromotor Operating environment conditions Temperature between 0 C and 28 C Pr...

Страница 22: ...gh speed handpieces Accessories and Detachable Parts Detachable Parts Optical fiber tip for curing light Curing light acrylic tip Light hood for curing light Ultrasonic scaler tips Ultrasonic scaler t...

Страница 23: ...23 11 Symbology The symbology complies with standards IEC 60601 1 and IEC 60878...

Страница 24: ...24 12 Dimensional Measurements in millimeters...

Страница 25: ...ns in chapter 9 Cleaning and Disinfection of this manual Before starting the office check the conditions of the compressor close the air reservoir drain and turn On it observe its operation until its...

Страница 26: ...g the procedure a Coupling or removal of the electrical micromotor from its coupling b Coupling or removal of the ultrasonic scaler from its coupling c Coupling or removal of high speed optical fiber...

Страница 27: ...Olsen certified assistance 3 Pneumatic handpiece has water leakage on the coupling 1 There s looseness on the handpiece coupling 1 Attach the handpiece correctly 2 The gasket is worn 2 Replace the ga...

Страница 28: ...the rotation s direction switch to FWD or REV position 4 The speed control is adjusted to minimum speed 4 Increase the micromotor s rotation speed 5 There s a pneumatic or electronic problem 5 Contact...

Страница 29: ...ith the warranty period Allow only qualified Olsen technicians to perform installation and maintenance on your equipment and accessories Use only Olsen original parts and accessories The use of non or...

Страница 30: ...Connection to wrong voltage power supply 5 This warranty will cease when a In the normal course of its period of warranty b Make changes in the product not authorized by Olsen c Adulterations in the p...

Страница 31: ...and of low maintenance cost These characteristics have been achieved thanks to our competent and dedicated team which make me very proud for many reasons as they are always giving the best of their cr...

Страница 32: ...________________________ __________________________________________________ __________________________________________________ __________________________________________________ ______________________...

Страница 33: ...________________________ __________________________________________________ __________________________________________________ __________________________________________________ ______________________...

Страница 34: ...__________________________________________________ __________________________________________________ __________________________________________________ _______________________________________________...

Страница 35: ...35...

Страница 36: ...36 55 48 2106 6000 export3 olsen odo br Technician in Charge Eng Cleber da Costa CREA SC S1 116283 5 Cod 5409249 Rev 03 22 03 2021...

Отзывы: