16
The Versatile Table needs special attention regarding electromagnetic compatibility and must be
installed and put into use in accordance with the electromagnetic compatibility information
presented in this chapter.
Radio frequency (RF) communication equipment, portable and mobile, can affect the Versatile
Table.
Guidance and Manufacturer's Declaration - Electromagnetic Emissions
The Versatile Table is intended for use in the electromagnetic environment specified below. The customer or the user should
assure that it is used in such an environment.
Emission test
Compliance
Electromagnetic Environment - Guidance
RF emissions
CISPR 11
Grupo 1
The Versatile Table uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to
cause any interference with nearby electronic equipment.
RF emissions
CISPR 11
Class B
The Versatile Table is suitable for use in all establishments,
including domestic establishments and those directly connected
to the public low-voltage power supply network that supplies
buildings use for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Non-applicable
Emissions due to voltage
fluctuations/scintillation
emissions
IEC 61000-3-3
Non-applicable
Guidance and Manufacturer's Declaration - Electromagnetic Emissions - I
The Versatile Table is intended for use in the electromagnetic environment specified below. The customer or the user should
assure that it is used in such an environment.
IMMUNITY Test
IEC 60601 Test Level
Compliance Level
Electromagnetic Environment Guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete or
ceramic file. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
± 2 kV for power supply
lines
±1 kV for input/output
lines
± 2 kV for power supply
lines
±1 kV for input/output
lines
Mains power quality should be that a
typical commercial or hospital
environment.
Surge
IEC 61000-4-5
± 1 kV differential mode
± 2 kV common mode
± 1 kV differential mode
± 2 kV common mode
Mains power quality should be that a
typical commercial or hospital
environment.
Voltage dips, short
interruptions and
voltage variations on
the power supply
input lines
IEC 61000-4-11
<5% U
T
(dip > 95% in U
T
)
for 0.5 cycle
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
<5% U
T
(dip > 95% in U
T
)
for 5 s
<5% U
T
(dip> 95% in U
T
)
for 0.5 cycle
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
<5% U
T
(dip> 95% in U
T
)
for 5 s
Mains power quality should be that of a
typical commercial or hospital
environment. If the user of the Versatile
Table requires continued operation during
power mains interruptions, it is
recommended that the Versatile Table be
powered from an uninterruptible power
supply or a battery.
Power frequency
(50/60 Hz) Magnetic
field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or
hospital environment.
NOTE U
T
is the a.c. mains voltage prior to applications of the test level.