19
- Curing Light (Optional)
Before starting the curing light use, install
the tip (1) and the light hood (2).
To install the tip: first, attach the light hood
(2) to the tip (1) and then attach it to the curing
light unit (3). The tip (1) must be pushed all the way to the fitting end.
To activate/deactivate the curing light: press the On/Off button (5) twice.
The green LED (4) indicates that the curing light is operating. The curing light unit
will run for 20 seconds and will turn Off automatically.
To toggle the operating mode: each tap on the selection button (9) switches
the operation mode. Press the selection button (9) until the orange LED indicates
the desired mode.
The curing light operates in the following modes:
−
Continuous (6): turns on the maximum power for 20 seconds;
−
Ramp (7): turns on the minimum power, gradually increasing. In 5
seconds reaches its maximum power until the cycle of 20 seconds is
finished;
−
Pulse (8): the device blinking for 20 seconds.
- Prophylaxis System
6.10.1
- Bicarbonate Jet
Before starting use of the bicarbonate jet,
its necessary deposit bicarbonate sodium in the
reservoir (4), check for moisture in the filter (8) and connect
the bicarbonate jet handpiece (3) to its coupling (2).
To fill the bicarbonate reservoir: open the reservoir (4) by
turning the cover (6) anticlockwise and deposit the bicarbonate
sodium until to the red line (5). Close the reservoir.
To remove condensed water in the filter: press the valve
(7) on the base of the filter (8) to drain the water.
To activate/deactivate the bicarbonate jet: press the
progressive/activation pedal (1). The more pressure on the pedal,
the greater the volume of the jet emitted by the handpiece, until
reaching the pressure limits of the equipment.
To adjust the spray: see
section 6.1
.
Clean the sodium bicarbonate tank (4) at the end of each procedure.
6.10.2
- Prophy-Jet
(Optional)
Before starting Prophy-Jet (3), deposit up to 15g of
sodium bicarbonate for dental use in the sodium bicarbonate
tank (4) and attach the Prophy-Jet to the pneumatic coupling
(2) available on the assistant module (
section 5.7
) or
pneumatic coupling for low-speed rotation in the working table
(
section 5.4
).
To activate/deactivate: push the propulsion/activation pedal (1). The higher the pressure applied to the
pedal, the greater the volume of the jet emitted by the handpiece, until reaching the pressure limits of the
equipment.
Clean the sodium bicarbonate tank (4) at the end of each procedure.
Содержание Infinity
Страница 1: ...1...
Страница 2: ...2...
Страница 23: ...23 8 Symbology The symbology complies with IEC 60601 1 IEC 60878 ISO 15223 1 ISO 7010...
Страница 27: ...27...
Страница 29: ...1...
Страница 30: ...2...
Страница 51: ...23 8 Simbolog a S mbolos en conformidad con las normas IEC 60601 1 IEC 60878 ISO 15223 1 ISO 7010...
Страница 55: ...27...