39
OLCT 200 User Manual |
39
Use the following step-by-step procedure to perform ZERO and SPAN calibrations.
1. To enter the CAL MODE from either data display, press the DOWN / CAL key
and within 5 seconds press the EDIT key.
Note
: During CAL MODE, the
follow signal is transmitted to the receiving device and signifies “CAL MODE”.
This special value is used to inhibit alarm trips at the receiver.
2 Wire Model
3 Wire Model
10-30VDC
Wireless Model
Battery Powered
Wireless Model
3mA
1.5mA
75 counts
(-15.6%FS)
75 counts
(-15.6%FS)
FOR OXYGEN (O2), SKIP STEPS 2 & 3 AND PROCEED DIRECTLY TO
THE SPAN
CALIBRATION BY PRESSING THE “NEXT” KEY.
2. Using the Cal-Cup, apply a clean ZERO gas or be sure there is no
background target gas in the monitored area. After the reading is stable,
(approximately 1 minute) press the EDIT key to perform a ZERO calibration.
3. If the ZERO calibration is successful, press the NEXT key to proceed to the
SPAN check.
4. Apply the
correct
SPAN gas at .5 liters/min. After the reading is stable,
(approximately 1 minute) press the EDIT key to perform a SPAN calibration.
WARNING:
The SPAN gas used must match the value specified since this is
what the OLCT 200 will indicate after a successful SPAN calibration. The
Cal
Span Value
may be edited if it becomes necessary to apply a different gas
concentration (see
Cal Span Value
in section 5.3).
5.
If the SPAN calibration is successful, the display flashes “REMOVE CAL
GAS” and starts the CAL PURGE delay.
Note
: During CAL PURGE delay,
transmitters are set to 0% full scale to avoid unintended alarm events.
Oxygen monitors are set to 20.9% oxygen during this delay.
6. CAL MODE will be complete after the end of the CAL PURGE delay.
The flow chart in Figure 4-2 illustrates the above procedure. UP, CAL, NEXT & EDIT
labels indicate keystrokes using the magnetic wand. The CAL MODE information
screen (top of the chart) is available for advanced users to see Offset / Gain calibration
constants and live analog to digital converter (A/D) counts. Span Gas calibration
values may also be edited from this screen. Holding the UP key, for 5 seconds during
CAL MODE, displays this screen.
Calibration history records are logged and may be viewed in the
Sensor Information
menu (see section 5-5).
Содержание OLCT 200
Страница 4: ...OLCT 200 User Manual 4...
Страница 14: ...OLCT 200 User Manual 14 Figure 2 3a Outline Drawing OLCT 200 Wired Models...
Страница 19: ...OLCT 200 User Manual 19 2 6 2 Installation Drawing 11 0100...
Страница 32: ...OLCT 200 User Manual 32...
Страница 36: ...OLCT 200 User Manual 36...
Страница 38: ...OLCT 200 User Manual 38 Figure 4 1 Calibration Gas Input...
Страница 61: ...OLCT 200 User Manual 61 Figure 6 5...
Страница 62: ...OLCT 200 User Manual 62...
Страница 79: ...OLCT 200 User Manual 79...