System Wiring
iTrans Fixed Point Single/Dual Gas Monitor
4-2
iTrans User Manual
(P/N: 77023554-1)
6.
Shielding from either the controller or remote sensors should be bonded to the
enclosure screw located inside the
.
IMPORTANT:
Use of this product in areas where it may be subject to large
amounts of electromagnetic interference may affect the reliable operation of this
device and should be avoided.
WARNING:
Supply wire with a minimum rating of 90
o
C must be used for
interconnection to the
.
NOTE:
For classified locations, a “poured” wire seal must be installed within
18 inches of the main unit for both power entry and remote sensors.
NOTE:
Remove power from the
before making any wiring connections.
4.3. Alarm Relay Wiring (J1, J5, and J6)
To connect the
control wires to the three relay terminals on the relay
board, wire the unit to the connectors shown in Figure 2-4.
The
low alarm relay
is activated when the low alarm threshold is met. This is a non-latching, Normally
Open (NO) contact. The
high alarm relay
is activated when the high alarm
threshold is met. This is a non-latching, Normally Open (NO) contact. The
fault
alarm relay
is activated upon power-up of the
. When the fault condition
is met, the circuit opens. This is an Electronically closed (NO) contact. See Figure
4-1 for relay wiring.
NOTE:
It is recommended that on-board relays should not be used to drive loads
directly. On-board relays should be used to drive a secondary, higher-power relay
which is connected to the control device (e.g., strobe, siren, exhaust fan, etc.).
Содержание iTrans
Страница 13: ...iTrans Fixed Point Single Dual Gas Monitor Introduction P N 77023554 1 iTrans User Manual 1 7...
Страница 14: ......
Страница 20: ...Hardware Overview iTrans Fixed Point Single Dual Gas Monitor 2 6 iTrans User Manual P N 77023554 1...
Страница 36: ......
Страница 54: ...Operation iTrans Fixed Point Single Dual Gas Monitor 5 18 iTrans User Manual P N 77023554 1...
Страница 62: ...ModBus Interface iTrans Fixed Point Single Dual Gas Monitor 6 8 iTrans User Manual P N 77023554 1...
Страница 68: ...Troubleshooting iTrans Fixed Point Single Dual Gas Monitor 8 4 iTrans User Manual P N 77023554 1...
Страница 78: ...Appendix D iTrans Fixed Point Single Dual Gas Monitor D 2 iTrans User Manual P N 77023554 1...
Страница 80: ...Appendix E iTrans Fixed Point Single Dual Gas Monitor E 2 iTrans User Manual P N 77023554 1...
Страница 86: ......
Страница 87: ......