Instructions for Use
03
Tested conformity with standards for emission and immunity
Emission
The device complies with the requirements for radiated and conducted RF-emissions according to CISPR 11,
Group 1, Class B.
Guidance and manufacturer’s declaration for electromagnetic compatibility
The OkuStim system is intended for the use in the electromagnetic environment specified below. The customer
or the user of the OkuStim system should assure that it is used in such an electromagnetic environment.
Immunity test
IEC 60601-test level
Compliance level
Electromagnetic
environment – guidance
Electrostatic discharge
(ESD according to IEC
61000-4-2)
±8 kV contact discharge
±2, 4, 8 und 15 kV air
discharge
±8 kV contact discharge
±2, 4, 8 und 15 kV air
discharge
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should
be at least 30 %.
Portable and mobile RF
communications equip-
ment should be used no
closer to any part of the
OkuStim system, inclu-
ding cables, than the
recommended separa-
tion distance calculated
from the equation appli-
cable to the frequency of
the transmitter.
Recommended separa-
tion distance:
Conducted RF-distur-
bances according to IEC
61000-4-6
3 V
Eff.
150 kHz to 80 MHz
outside ISM and amateur
bands
3 V
Eff.
150 kHz bis 80 MHz
outside ISM and amateur
bands
d=1,2√P
Radiated RF-disturban-
ces according to IEC
61000-4-3
10 V/m
80 MHz to 2,7 GHz
10 V/m
80 MHz to 2,7 GHz
Limit values for RF
communication devices
according to table 9
in IEC 60601-1-2
(9- 28 V/m)
Rev. 2021-07-01
OkuStim
System
TD60K06 -
33 36
Содержание OkuStim
Страница 1: ...Instructions for Use 03 Instructions for Use EN Rev 2021 07 01 OkuStim System TD60K06 OkuStim System...
Страница 34: ...Instructions for Use 03 Notes Rev 2021 07 01 OkuStim System TD60K06 34 36...
Страница 35: ...Instructions for Use 03 Notes Rev 2021 07 01 OkuStim System TD60K06 35 36...
Страница 36: ...Instructions for Use 03 Okuvision GmbH 2021 Rev 2021 07 01 OkuStim System TD60K06 36 36...