70
Figure 95. CO2 Zero Reset procedure with Air completed.
If the
CO2 Zero Reset
procedure has failed the message shown in Figure 96 appears. In this case Okolab
recommends to repeat the
CO2 Zero Reset
procedure and if it fails again please contact Okolab support at
.
Figure 96. CO2 Zero Reset procedure with Air has failed.
13.4.2
CO2 Zero Reset with N2
To start the
CO2 Zero Reset
procedure using N2, press on
N2
as shown in Figure 97 a. Pressing on
N2,
LEO
reminds you to connect a Nitrogen bottle at 1.0 barg (14.5 psig) to LEO. See Figure 97 b.