H301-NIKON-NZ100/200/500-N
User Manual
Rev 03
www.oko-lab.com
Pa
ge
5
3.
Available Lids
The following heated glass Lids are available for H301-NIKON-NZ100/200/500-N:
•
H301-SLIDING-LID
: For easy sample loading and pipetting
–
Included in code
H301-NIKON-NZ100/200/500-N.
Compatible with Mad City Labs NZ100/200 ONLY.
•
H301-KOEHLER-LID*
: Reduces chamber height to 22 mm and allows imaging under Koehler illumination
•
H301-INJECTION-LID*
: Heated glass lid with two small openings (sealed with flexible plastic) allowing injec-
tion or permanent access to the sample. Compatible only with sample holder 1x35
•
H301-LASER-INTERLOCK-LID*
: Heated glass lid with safety switch. Connects to laser controller and automati-
cally turns laser off when lid is lifted. Compatible only with sample holder 1x35
* OPTIONAL
–
not included with H301-NIKON-NZ100/200/500-N
3.1
Sliding lid
The Sliding Lid is a heated glass lid allowing for easy sample loading and pipetting. The Sliding Lid is screwed onto the
chamber.
NOTE:H301-SLIDING-LID is compatible with Mad City Labs NZ100/200 ONLY.
Figure 3 shows chamber dimensions with the Sliding Lid (with and without chamber riser). Figure 4 illustrates how to
remove Sliding Lid, when a different Lid is necessary.
NOTE:
Sliding Lid MUST BE REMOVED when using any other Lid.
The Sliding Lid is fixed onto the chamber with 4
screws. Screws location is indicated by letter A in Figure 4 (image 2 and 3). Keep Sliding Lid fully closed to access
screws labeled A in image 2. Open Sliding Lid to access screws labeled A in image 3.
Distance between focal plane
and upper side of the glass
lid without the chamber riser
Distance between focal plane
and upper side of the glass
lid with the chamber riser
Chamber top view
Sliding Lid
Figure 3. H301-NIKON-NZ100/200/500-N (Sliding Lid comes as standard)