Connecting to a USB port    If your 

computer has USB ports you can get 

an optional RS-232 to USB adapter 

from Ohmic. Ohmic recommends 

using an RS-232 to USB adapter only 

with Windows XP

©

 and Windows 

Vista

©

. Connect the AMM-15 using 

the RS-232 adapter and RS-232 to 

USB adapter. Turn on your computer. 

When your computer recognizes the 

device click OK and follow the instruc-

tions for installing the driver software. 

Set up your communications program 

as described below. 

 

Communications Programs 

 

The AMM-15 transmits data via its RS
-

232 output as soon as it is powered 

up. In order to receive the data, a 

communications (or terminal) program 

must be running on your computer. 

We recommend 

Tera Term;

 it is easy 

to use, free, and available for down-

load from www.download.cnet.com. 

When you reach that page, type “Tera 

Term” in the Search box; you will then 

be able to download the installation 

program. Then: 
 

1.  Install the program and run it. 

2.  A 

New Connection

 window will 

open. Click the “Serial” button, 

then select the COM port you will 

be using, and click OK. 

3.  Tera Term should now be ready to 

receive data from the AMM-15. 

Make sure they are connected via 

the adapter cable, and power up 

the AMM-15. 

4.  Text will be displayed indicating 

the version of the embedded pro-

gram. The computer will then dis-

play four numbers, as indicated 

below:  

 

(Date)  (Time)  (Dew Point) (Temp.)

 

mm/dd/yy   hh/mm/ss          xx.x                xxx.x 

 

 

 

New readings will be displayed 

every few seconds and are in a 

comma delimited format that can 

be copied and pasted into a 

spreadsheet program, if desired. 

If readings are not displayed,

 turn the 

AMM-15 off, and in Tera Term click 

on 

Setup

 in the menu bar, then se-

lect 

Serial Port

. This will open a win-

dow where you can check several 

parameters: 
 

• 

Port: 

Try another port until you 

find the right one. 

• 

Baud rate: 

Should be 9600. 

• 

Data: 

8 bit 

• 

Parity: 

None 

• 

Stop: 

1 bit 

• 

Flow Control: 

Off  

 

Once you get a working display you 

may want to save your settings. Re-

fer to Tera Term’s help file for infor-

mation on this and other functions. 
 

Other communications programs can 

also be used; some older versions of 

Windows include one called Hyper-

Terminal. Refer to its help file for 

information on using it and applying 

the parameters above. 

Содержание AMM-15

Страница 1: ...AMM 15 Phone 410 820 5111 Toll Free 800 626 7713 Fax 410 822 9633 www ohmicinstruments com Sales sales ohmicinstruments com Service service ohmicinstruments com Ohmic Instruments 3081 Elm Point Indust...

Страница 2: ...may be copied as required by the original purchaser of the instrument Information contained in this manual is believed to be accurate and reliable however no responsibility is assumed by Ohmic Instrum...

Страница 3: ...formation 6 Operating Instructions 8 RS 232 Interface 8 Logging and Data Uploading 10 Changing Settings Using the Interior DIP Switch 10 Summary of Computer Commands 11 Notes on Communication and Comm...

Страница 4: ...he sensor elements the unit should be stored at 0 to 125 Deg F at less than 90 RH Long term storage of the AMM 15 out of these conditions is not recommended by Ohmic Instruments We also recommend remo...

Страница 5: ...humidity and dry bulb temperature of the sample air from the system under test Dew point is then calculated from these measurements using pre programmed psychrometric equations and is fully compensate...

Страница 6: ...pressure is insuffi cient to provide positive flow a vacu um pump can be connected to the second sample port to help draw a sample Monitoring Compressed Air or Pressurized Vessels In compressed air ap...

Страница 7: ...ITH 5 PSI OR GREATER This orifice maintains the test cell at line pressure if the orifice is omitted false readings will result Filter Brass Orifice Assembly Outlet Inlet Hose barb to Quick Connect OD...

Страница 8: ...int and an up arrow when displaying Dry Bulb Temperature The F C selector button will change the displayed parameter from degrees Fahren heit to degrees Celsius 5 Allow sufficient time for the read in...

Страница 9: ...ect the COM port you will be using and click OK 3 Tera Term should now be ready to receive data from the AMM 15 Make sure they are connected via the adapter cable and power up the AMM 15 4 Text will b...

Страница 10: ...rolling speed of the terminal screen might stretch this interval to about four and a half minutes Uploading will not clear the logger memory which makes repeated up loading of the same data possible D...

Страница 11: ...1 12 DD day 01 31 YY year 00 99 hh hours 00 23 mm minutes 00 59 ss seconds 00 59 Clear stored data Upload stored data Readings including a time stamp are transmitted in chronological order one line co...

Страница 12: ...B CR LF where the numbers corre spond to Dew Point and Tem perature respectively Each scan of the sensors is preceded by a time stamp in the form MM DD YY hh mm ss For example 04 10 13 16 30 05 is Apr...

Страница 13: ...00 73 853 33 40 100 00 2 353 33 36 606 67 74 860 00 39 106 67 1 360 00 37 613 33 75 866 67 38 113 33 0 366 67 38 620 00 76 873 33 37 120 00 1 373 33 39 626 67 77 880 00 36 126 67 2 380 00 40 633 33 7...

Страница 14: ...racy 2 F Average Response Time 10 Seconds 63 Step Max Inlet Temperature 140 F 550 F with Heat Sink Tubing Maximum Line Pressure 120 PSIG Display 3 1 2 Digit LCD 0 5 Character Height Power 4 AA Alkalin...

Страница 15: ...ed transpor tation prepaid SHIPPING TO MANUFACTURER FOR REPAIR OR ADJUSTMENT All shipments of Ohmic Instruments products should be made via United Parcel Service or Best Way prepaid The instrument sho...

Страница 16: ...16...

Отзывы: