OPERATING INSTRUCTIONS 

1.  Make sure the alkaline batteries 

have sufficient power. The display 

will show the low battery symbol if 

the batteries are inadequate. If the 

instrument has not been used for 

a period of time, the batteries 

should be checked and replaced if 

necessary. If desired, the AMM-15 

can be operated from the AC 

Adapter (it does not charge the 

batteries). As with all such equip-

ment, if the unit is not to be used 

for an extended period, the batter-

ies should be removed. 

2.  Purge the air line to be tested of 

any water, oil, or other foreign 

matter. The sensors are suscepti-

ble to contamination and damage 

from impurities such as water, oil, 

dust, chemicals, etc.  

THE PRO-

VIDED FILTER MUST BE BE-

TWEEN THE AMM-15 AND THE 

SYSTEM BEING SAMPLED 

FROM.  FAILURE TO USE THE 

FILTER VOIDS THE WARRAN-

TY.  OHMIC INSTRUMENTS IS 

NOT RESPONSIBLE FOR MIS-

USE OR NEGLECT OF THE 

UNIT.

 

3.  Connect the AMM-15 to the air 

line following the recommenda-

tions on the  Sample Modes / 

Connecting Information page.  

[AIR LINE PRESSURE IS NOT 

TO EXCEED 120 PSI] 

4.  Press the ON button. The AMM-

15 always displays Dewpoint on 

power up.  The DP / DB selector 

button will change the displayed 

parameter from Dewpoint to Dry 

Bulb Temperature. The display 

will show a "down" arrow when 

displaying Dewpoint and an "up" 

arrow when displaying Dry Bulb 

Temperature. The °F / °C selector 

button will change the displayed 

parameter from degrees Fahren-

heit to degrees Celsius. 

5.  Allow sufficient time for the read-

ings to stabilize. Normal stabiliza-

tion will occur within 1 to 1-½ 

minutes. Very low readings, less 

than -10°F DP, will take up to 10 

or 15 minutes for stabilization.  

Analog Output 

The AMM-15 provides an analog 

output of 0-1 VDC (from –55 to +95°

F Dew Point). Use the enclosed 2-

wire plug adapter, and connect a 

voltmeter to the wires:  Red/white to 

positive (+) and black/white to nega-

tive (-). Plug the adapter into the jack 

marked Analog Out, and see the 

table on Page 13 for values. 

RS-232 INTERFACE– Connect-

ing the AMM-15 to your Comput-

er 

 

Connecting to a serial port   The 

AMM-15 can be plugged into any 9-

pin PC serial (COM) port using a 9-

pin serial cable, available from most 

computer stores. If your computer 

has a 25-pin COM port, 25-pin to 9-

pin serial adapters are also available 

from most computer stores and can 

be used with the cable [Serial data is 

transmitted using Pin 5 (ground), Pin 

3 (transmit), and Pin 2 (receive)]. 

Some computer COM ports and seri-

al adapters are not fully implemented 

(not all pins are used). If the AMM-15 

is connected to a partially implement-

ed serial port, functions may be errat-

ic or the unit may fail to communi-

cate. Please consult the owner’s 

manual for your computer if this oc-

curs. Set up your communications 

program as described on the next 

page. 

Содержание AMM-15

Страница 1: ...AMM 15 Phone 410 820 5111 Toll Free 800 626 7713 Fax 410 822 9633 www ohmicinstruments com Sales sales ohmicinstruments com Service service ohmicinstruments com Ohmic Instruments 3081 Elm Point Indust...

Страница 2: ...may be copied as required by the original purchaser of the instrument Information contained in this manual is believed to be accurate and reliable however no responsibility is assumed by Ohmic Instrum...

Страница 3: ...formation 6 Operating Instructions 8 RS 232 Interface 8 Logging and Data Uploading 10 Changing Settings Using the Interior DIP Switch 10 Summary of Computer Commands 11 Notes on Communication and Comm...

Страница 4: ...he sensor elements the unit should be stored at 0 to 125 Deg F at less than 90 RH Long term storage of the AMM 15 out of these conditions is not recommended by Ohmic Instruments We also recommend remo...

Страница 5: ...humidity and dry bulb temperature of the sample air from the system under test Dew point is then calculated from these measurements using pre programmed psychrometric equations and is fully compensate...

Страница 6: ...pressure is insuffi cient to provide positive flow a vacu um pump can be connected to the second sample port to help draw a sample Monitoring Compressed Air or Pressurized Vessels In compressed air ap...

Страница 7: ...ITH 5 PSI OR GREATER This orifice maintains the test cell at line pressure if the orifice is omitted false readings will result Filter Brass Orifice Assembly Outlet Inlet Hose barb to Quick Connect OD...

Страница 8: ...int and an up arrow when displaying Dry Bulb Temperature The F C selector button will change the displayed parameter from degrees Fahren heit to degrees Celsius 5 Allow sufficient time for the read in...

Страница 9: ...ect the COM port you will be using and click OK 3 Tera Term should now be ready to receive data from the AMM 15 Make sure they are connected via the adapter cable and power up the AMM 15 4 Text will b...

Страница 10: ...rolling speed of the terminal screen might stretch this interval to about four and a half minutes Uploading will not clear the logger memory which makes repeated up loading of the same data possible D...

Страница 11: ...1 12 DD day 01 31 YY year 00 99 hh hours 00 23 mm minutes 00 59 ss seconds 00 59 Clear stored data Upload stored data Readings including a time stamp are transmitted in chronological order one line co...

Страница 12: ...B CR LF where the numbers corre spond to Dew Point and Tem perature respectively Each scan of the sensors is preceded by a time stamp in the form MM DD YY hh mm ss For example 04 10 13 16 30 05 is Apr...

Страница 13: ...00 73 853 33 40 100 00 2 353 33 36 606 67 74 860 00 39 106 67 1 360 00 37 613 33 75 866 67 38 113 33 0 366 67 38 620 00 76 873 33 37 120 00 1 373 33 39 626 67 77 880 00 36 126 67 2 380 00 40 633 33 7...

Страница 14: ...racy 2 F Average Response Time 10 Seconds 63 Step Max Inlet Temperature 140 F 550 F with Heat Sink Tubing Maximum Line Pressure 120 PSIG Display 3 1 2 Digit LCD 0 5 Character Height Power 4 AA Alkalin...

Страница 15: ...ed transpor tation prepaid SHIPPING TO MANUFACTURER FOR REPAIR OR ADJUSTMENT All shipments of Ohmic Instruments products should be made via United Parcel Service or Best Way prepaid The instrument sho...

Страница 16: ...16...

Отзывы: