Страница 1: ...t Halogen Phototherapy Light II Operation and Maintenance Manual i i 1 i ii i iii Uk CO O B XOO iW de I boS 0 HHT M3 0 3 L POL v...

Страница 2: ...76 microwatts square centimeter nanometer in the 425 475 nanometer spectrum at bed level depending on lamp to patient distance Lamp intensity 6 500 foot candles Lamp spot size Variable from 3 to 20 i...

Страница 3: ...s satisfactory when properly installed on a typical wood frame drywall However different hardware may be required for other wall constructions 2 1 INSTALLATION ON THE INFANT WARMER SYSTEM WARNING Thes...

Страница 4: ...oles in the backplate and the transformer housing Fig ure 2 Secure the transformer housing and backplate to the mounting bracket Use two nuts as shown in Figure 2 Tighten the nuts Mounting Bracket Sli...

Страница 5: ...on arm and the articulating arm This washer is required for proper arm operation g Insert the brass friction plug concave surface first into the hole in the rear of the articulating arm Insert the spr...

Страница 6: ...tes to the transformer housing Peel the paper backing off of the adhesive strips and fit the cover plates into the recesses on the top and bottom of the transformer housing Figure 2 10 Replace the end...

Страница 7: ...the lamp head and the articulating arm Figure 4 Figure 4 Tension Joint b Rotate the lamp head counterclockwise until the notch lines up with a hex head screw at the top of the joint c Turn the screw...

Страница 8: ...c drill with 1 4 x 1 1 2 drill bit 3 16 inch hex key wrench included 9 16 inch nut driver or socket wrench Medium size flat blade screwdriver Mallet 1 Locate a wood stud in the wall near the desired m...

Страница 9: ...Washer I pivot Cylinder Wire Cover Panel Lag Screws Cover Plats Adhesive Strips Transforms Housing Wood Stud Power Cord Figure 5 Wall Mounting the Phototherapy Light...

Страница 10: ...Articulating Arm Lamp head f IfttT Bod Cap Caster Mounting Figure 6 Assembling The Floor Stand Phototherapy Light page 15 Detail of...

Страница 11: ...securely to the lens housing If necessary tighten the set screw on the underside of the lens bezel 6 For floor stand models check that the pole assembly is securely mounted to the base Tighten the mo...

Страница 12: ...If the unit fails any part of the checkout procedure it must be repaired and tested before use on a patient...

Страница 13: ...r of the lamp head assembly during operation 3 1 CONTROLS Figure 7 A On Off Switch On the rail wall mounting model the On Off switch is located on the column at the top of the transformer housing On t...

Страница 14: ...Figure 7a Phototherapy Light Controls Rail Wall Mounted Model Figure 7b Phototherapy Light Controls Floor Stand Model...

Страница 15: ...the largest of the four available aperture sizes for phototherapy WARNING Prolonged exposure to the ultraviolet radiation emitted by this light may harm the unprotected eyes of the infant or operator...

Страница 16: ...s perform the checkout procedure before using the equipment on a patient Repair and service by an authorized Ohmeda Service Representative insures full reliability If Ohmeda Service is not available p...

Страница 17: ...rest articles or liquids on top of the Phototherapy Light Liquids will damage the lamp head and power supply assemblies causing an electrical shock hazard CAUTION Use only lens cleaner or a similar no...

Страница 18: ...Open the top cover tilt the head to allow the cover to open fully past the yoke 5 The lamp shipped with the unit is wired into the holder to prevent movement during shipment Remove this copper wire an...

Страница 19: ...N ADJUSTMENT WARNING Disconnect the Phototherapy Licfht power cord and allow the unit to cool for at least 30 minutes before servicing The Phototherapy Light operates at a high temperature Adjust the...

Страница 20: ...hex key wrench included with the unit Check that the lamp head assembly does not move more than one inch after being positioned 5 Complete the electrical safety check the checkout procedure and a leak...

Страница 21: ...plete the electrical safety check the checkout procedure and a leakage current test before using the Phototherapy Light on a patient If the fuse continues to open the unit should be removed from use a...

Страница 22: ...thereof Suggestions are invited from our customers for consideration by Ohmeda with these periodic reviews WARNING After completing a repair of the Phototherapy Light perform the checkout procedure to...

Страница 23: ...RNING Disconnect the Phototherapy Light power cord and allow the unit to cool for at least 30 minutes before servicing The Phototherapy Light operates at a high temperature CAUTION Be careful not to t...

Страница 24: ...y help to wiggle the socket slightly while pushing 13 Slide the lamp all the way down into the lamp bracket and under the spring wires 14 Close the top cover and fully tighten the slotted screw 15 Com...

Страница 25: ...ade from the old motor shaft 13 Mount the fan blade on the new motor The collar should face away from the motor toward the vents 14 Install the new motor in position and mount it with the two mounting...

Страница 26: ...minutes 2 Use a small screwdriver and carefully lever the power switch out of the housing 3 Unsolder the wire connections and discard the switch 4 Slip heatshrink tubing over the wires 5 Solder the wi...

Страница 27: ...the backplate 6 Remove the four screws lock washers and flat washers securing the transformer to the transformer housing a Also loosen and remove the screw securing the green wire ground to the transf...

Страница 28: ...rom the lamp nead assembly b Molex connector I 1 Crimp the blue and black transformer primary wires together with one Molex pin 2 Insert the blue and black transformer primary wires into pin 2 13 Mate...

Страница 29: ...the unpainted cylinder into the articulating arm and replace the wire cover panel on the underside of the extension arm 20 Remount the Phototherapy Light assembly as detailed in section 2 1 rail and...

Страница 30: ...sulation on the ends of the new transformer wires v 9 Use wire nuts to connect the new transformer wires as follows Reference Section 7 Schematics a Crimp the white power cord wire and the yellow and...

Страница 31: ...y a Position the lamp head assembly near the pole and connect the three connectors b Slide the lamp head assembly into the pole assembly c Rotate the assembly until it is properly centered over the tr...

Страница 32: ...60Qi 6 Threaded plug 7 Plastic washer Vizz 8 Transformer housing VznQ 9 Wire cover panel Vyi 10 Cover panel screw Q0 11 End cap arm 6600 12 Cover plate 6600 13 Lag screw 6600 14 Filtering lens 600 15...

Страница 33: ...bracket 6600 0095 400 Transformer 6600 0024 200 Screw transformer mounting 6600 0111 200 Two pin female Molex connector 0690 1565 308 Female pin 0 062 0690 2600 343 Two pin male Molex connector 0690...

Страница 34: ...Assembly Internal Parts Description Stock Number c Olt 0 co 1 LamD 6600 6HH 6J3 9 2 S socke 6600 i0nn 3 Fan Motor 660 200 4 Fan hlade 6600 0025 200 Filtering UAs 6600 0026 200 6 Aperture wheel 6600 0...

Страница 35: ...dle assembly 6600 0035 200 4 Spring 6600 0102 200 5 Threaded plug 6600 0103 200 6 Base assembly 6600 0113 200 7 Plastic washer 6600 0105 200 8 Pole 6600 0114 200 9 Pivot shaft 6600 0115 200 10 8 32 se...

Страница 36: ...19 200 Screw transformer mounting 6600 0120 200 Screw transformer mounting plate 6600 0121 200 5 16 nut 6600 0119 200 Two pin female Molex connector 0690 1565 308 Female pin 0 062 0690 2600 343 Two pi...

Страница 37: ...II SCHEMATICS WALL MOUNT CASTING i EXTENSION ARM CONNECTORS Tf T J J PIVOT CONTROL ASSY 115V 60 Hz SUPPLY Figure 15 Wall Rail Mount Model V k...

Страница 38: ...r SSV 11SV 60 Hz SUPPLY Figure 16 Floor Stand Model PIVOT CONTROL ASSY UPRIGHT CONNECTORS BASE...

Страница 39: ...the expendable parts The foregoing warranties shall not apply if the Product has been repaired other than by Ohmeda or in accordance with written instructions provided by Ohmeda or altered by anyone...

Страница 40: ...A Southeastern Region 4565 Winters Chapel Rd Atlanta GA 30360 800 241 4300 Outside GA 404 449 7990 In GA Southwestern Region 831 Greenview Drive Grand Praine TX 75050 800 433 5070 Outside TX 800 772 5...

Страница 41: ...ion 2 Specifications No changesrequired Section 3 Setup and Checkout Procedure No changes required Section 4 CaIibration and Adjustments No changes required Section 5 Disassembly and Repair Replace pa...

Страница 42: ...nfant Warmer System to and from the floor for inspection or repair use two people for safety Always check to ensure that you lay the unit on its right side as viewed from the front when laying the uni...

Страница 43: ...ge area 5 Remove the hydraulic system assembly for replacement Mounting Screw Outer Cover Plato Inner Cover Plate 19 X 26 BED Bed Plotfor Roaove MountIng Nut Fro Stud Figure 5 13 Hydraulic System Asse...

Страница 44: ...0 0013 300 7 Bed Tilt Lever 0217 5277 100 8 Pin 0217 5321 500 9 Hydraulic System Assembly 0217 5360 810 Ball Joint 0217 5318 300 10 Valve Bed Hydraulic Plate 0217 5283 100 11 Bed Model 5000 0217 5381...

Страница 45: ...277 100 9 Pin 0217 5321 500 10 Hydraulic System Assembly 0217 5360 810 Ball Joint 0217 5318 300 11 Valve Bed Hydraulic Plate 0217 5283 100 12 Bed Model 3300 0217 5316 100 13 Pin 0217 5322 500 14 Pivot...

Страница 46: ...0203 3262 300 4 Support Button 0217 5275 100 5 End Bracket 0217 5291 100 6 Mounting Screw 0144 2130 224 Figure 7 16 WindowAssembly 19 x 26 BED Description Part Number 1 Window Front and Rear 6600 002...

Страница 47: ...sher 0144 1110 131 4 Left and Right Bed Side 25 In 6600 0009 700 5 Front and Rear BedSide 18 In 6600 0008 700 Figure 7 15 Bed Assembly 19 x 26 BED Description part Number 1 Foam Matress 0305 5060 300...

Отзывы: