EN-28

3

000 Series Indicators

5    SERIAL COMMUNICATION

 

The Indicators include an RS232 serial communication interface.  An optional RS232 serial communication interface can also  

be installed if required.

Note:

 Some indicators does not include an RS232 serial communication interface. An optional RS232 serial communication 

interface (PN:30101019) needs to be purchased.

The setup of RS232 operating parameters are more fully explained in Section 3.8.  The physical hardware connection is 
explained in in Section 2.2.

5.1     Interface Commands

The interface enables display data to be sent to a computer or printer.  A computer can be used to control some functions of 
the indicator using the commands listed in Table 5-1.

Command 
Character

 

 

Legacy
Command
(2) 

 

Function 

 

IP

Immediate Print of displayed weight (stable or unstable).

P

Print stable displayed weight (according to stability setting).

CP

CA

Continuous Print.

SP

Print when stable.

xS

0S: Turn off “Stable Only” menu item and allow unstable print.  1S: Turn on “Stable Only” menu 

item and only print stable print.

xP

xA

Interval Print x = Print Interval (1-3600 sec), 0P turns auto print OFF

Z

Same as pressing Zero button 

T

Same as pressing Tare button 

xT

Download Tare value in grams (positive values only).  Sending 0T clears tare (if allowed)

PU

Print current unit: g, kg, lb, PCS

xU

Set scale to unit x: 1=g, 2=kg, 3=lb

xM

Set scale to mode x. M will scroll to next available mode.

PV

V

Version: print name, software revision and LFT ON (if LFT is set ON).

Esc R

Global reset to reset all menu settings to the original factory defaults

NOTES

:

  • Commands sent to the Indicator must be terminated with a carriage return-line feed (CRLF). 
  • Data output by the Indicator is always terminated with a carriage return-line feed (CRLF).
  • The Legacy Commands maintain compatibility with older products

5.2     Output Format

Definitions:  

      

Weight

, up to 11 characters, right justified, ‘-‘ at immediate left of most significant character (if negative).

      

Unit

, up to 5 characters, right justified. If the Unit in the Print Content menu was set to OFF, the unit will be removed    

          in the weight string, and 5 spaces will be printed.

      Stability

, “?” character is printed if not stable, 1 space if stable.

      

G/N/T

: “N” printed if weight is net weight, ‘G’ or space printed if weight is a gross weight.

      

Terminating Character(s)

 – terminating character(s) printed depending on FEED menu setting.

The default serial output format is shown below.

Field:

Weight Space

Unit Space Stability

Space

G/N/T

Term. Char(s)

Length:

11

1

5

1

1

1

1

Содержание TD32PE

Страница 1: ...i 3000 Series Indicators Instruction Manual TD32PE Indicator...

Страница 2: ...ii...

Страница 3: ...icator EN 10 2 2 4 RS232 Interface Cable to Indicator Optional EN 10 2 3 Internal Connections EN 11 2 3 1 Opening the Housing EN 11 2 3 2 Jumper Connections EN 11 3 SETTINGS EN 12 3 1 Menu Structure E...

Страница 4: ...out EN 20 3 6 Mode Menu EN 21 3 6 1 Reset EN 21 3 6 2 Parts Counting Mode EN 21 3 6 3 End Mode EN 21 3 7 Unit Menu EN 21 3 7 1 Reset EN 21 3 7 2 Kilogram Unit EN 21 3 7 3 Gram Unit EN 21 3 7 4 Pound U...

Страница 5: ...6 1 Weighing EN 26 4 6 2 Parts Counting EN 26 4 6 3 Totalization EN 27 5 SERIAL COMMUNICATION EN 28 5 1 Interface Commands EN 28 5 2 Output Format EN 28 5 3 Printout Examples EN 29 6 LEGAL FOR TRADE E...

Страница 6: ...ut the product Warning Symbols General Hazard Electrical Shock Hazard 1 2 Safety Precautions Verify that the AC adapter s input voltage range and plug type are compatible with the local AC power to be...

Страница 7: ...TD32PE Indicator Item Description 1 Front Housing 2 Control Panel 3 Power Receptacle 4 Load Cell Cable Connector 5 RS232 Connector optional 6 Data Label 7 FCC Information 8 Rear Housing 9 Mounting Tra...

Страница 8: ...tion 1 Power Connector J1 2 Battery Connector J9 3 Option Connector J2 4 LFT Switch 5 4 6 Lines Sense Jumper W1 located on the other side of PCB 6 Load Cell Connector J4 7 4 6 Lines Sense Jumper W2 lo...

Страница 9: ...Gram symbols 2 Battery function symbol 3 TARE Menu button 4 FUNCTION Mode button 5 Accumulation Symbol 6 NET function symbol 7 PRINT Units button 8 PCS function symbol 9 Stable weight Symbol 10 ON ZER...

Страница 10: ...ng Unit Mode Allows changing the application mode Press and hold allows scrolling through modes Menu Enter the User menu Menu Function Short Press Yes Accepts the current setting on the display No Adv...

Страница 11: ...2 1 Scale Base to Indicator Connect the load cell cable to the indicator as shown below 2 1 Note To connect TD32PE to other bases a loadcell connector needs to be soldered as shown above 2 2 2 AC Powe...

Страница 12: ...e will disappear The indicator can operate for up to 80 hours on a fully charged battery During battery operation a flashing triangle above the battery function symbol indicates the battery is low and...

Страница 13: ...includes sense wires Jumpers W1 and W2 must be opened Note 6 wire load cell setting is default After wiring is completed and jumpers are in place replace the indicator housing screws Figure 2 2 Jumpe...

Страница 14: ...N ON PARITY 0 001 20 2d TOTALIZE GRAM 7 EVEN POWER ON UNIT 5d OFF OFF 7 ODD AUTO FILTER ON ON 7 NONE GRAM LOW END MODE LB 8 NONE KILOGRAM MED OFF STOP ZERO RANGE HI ON 1 0 AZT END UNIT 2 2 OFF HANDSHA...

Страница 15: ...Allows entry into the displayed menu Accepts the displayed setting and advances to the next menu item No Skips by the displayed menu Rejects the displayed setting or menu item and advances to the nex...

Страница 16: ...tion Linearity calibration uses 3 calibration points The first calibration point is established with no weight on the scale The second calibration point is established at approximately half capacity T...

Страница 17: ...gravity Attention Changing the GEO Factor alters the calibration The GEO value was set at the factory and should only be changed by an authorized manufacturer s representative or certified veriricati...

Страница 18: ...4 13 13 12 12 11 11 10 40 05 41 12 16 15 15 14 14 13 13 12 12 11 11 41 12 42 19 16 16 15 15 14 14 13 13 12 12 11 42 19 43 26 17 16 16 15 15 14 14 13 13 12 12 43 26 44 32 17 17 16 16 15 15 14 14 13 13...

Страница 19: ...used for the first time enter this menu to set the Capacity and Graduation 3 4 1 Reset Reset the Setup menu to the factory defaults No not reset Yes reset NOTE If the Legal for Trade switch is switche...

Страница 20: ...20 25 30 40 50 60 75 100 120 150 200 250 300 400 500 600 750 1000 1200 2500 0 1 0 2 0 5 1 5 10 15 20 25 30 40 50 60 75 100 120 150 200 250 300 400 500 600 750 1000 1200 1500 3000 0 1 0 2 0 5 1 5 10 15...

Страница 21: ...Range 0 5d 1d 2d 5d Filter Lo Med Hi Auto Zero Tracking Off 0 5d 1d 3d Sleep Off On Light HI Med Low Auto Off Off 1 5 10 min Expand Off On End Readout Exit READOUT menu 3 5 1 Reset Set the Readout men...

Страница 22: ...minal sleep functionality OFF Terminal will not sleep ON Terminal will sleep 3 5 7 Auto Off Set the automatic shut off functionality OFF disabled 1 powers off after 1 minute of no activity 5 powers of...

Страница 23: ...the desired units Default settings are bold 3 7 1 Reset Set the Unit menu to the factory defaults NO not reset YES reset Reset No Yes Kilograms Off On Grams Off On Pounds Off On End Unit Exit UNIT men...

Страница 24: ...ple Single Feed Line Feed 4 Line Feed Form Feed End Print Exit PRINT menu 3 8 1 Reset Set the Print menu to factory defaults NO not reset YES reset a 3 8 Print Menu appears only with RS232 option inst...

Страница 25: ...mediately ON values are only printed when the stability criteria are met 3 8 4 Stop Bit Set the number of stop bits 1 1 stop bit 2 2 stop bits 3 8 8 Content Select the additional content of the printo...

Страница 26: ...ayout criteria FORMAT Multi mutliple lines are printed Single single line is printed FEED Line move paper up one line after printing 4 Lines move paper up four lines after printing Form move paper to...

Страница 27: ...urn the Indicator off press and hold the ON ZERO Off button until OFF is displayed 4 2 Zero Operation Zero can be set under the following conditions Automatically at Power On initial zero Semi automat...

Страница 28: ...or sending the data to a computer requires that the communication parameters in the Print Menu are set refer to Section 3 8 Press the PRINT Units button to send the displayed data to the communication...

Страница 29: ...Totalization or adds the new displayed value to the accumulated total and displays the new total value The Accum indicator will blink when the value is shown Display accumulated data To display the a...

Страница 30: ...Zero button T Same as pressing Tare button xT Download Tare value in grams positive values only Sending 0T clears tare if allowed PU Print current unit g kg lb PCS xU Set scale to unit x 1 g 2 kg 3 lb...

Страница 31: ...4 KG G Gross value line If Print Content Gross is ON and a tare value is entered 11 11 KG N Net value line If Print Content Net is ON and a tare value is entered 1 22 KG T Tare value line If Print Con...

Страница 32: ...ing menu settings cannot be changed Span Calibration Linearity Calibration Calibration Unit GEO Capacity Graduation Zero Range Stable Range AZT Modes Units To enable editing of these menu settings ret...

Страница 33: ...ng PCS display Busy message Displayed during tare setting zero setting printing NO Action not allowed message Function not executed Battery icon flashing Low Battery error Battery is empty CAL E Calib...

Страница 34: ...2 Operating Temperature Range 10 C to 40 C 14 F to 104 F Housing Dimensions W x D x H 210 x 168 x 80 mm 8 27 x 6 61 x 3 15 in Net Weight 1 4 kg 3 1 lb Shipping Weight 2 4 kg 5 3 lb Shipping Dimensions...

Страница 35: ...EN 33 3000 Series Indicators 8 3 Drawings and Dimensions Figure 8 1 Indicator Overall Dimensions 3 28 83 5 mm...

Страница 36: ...l interference to radio communications Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at...

Страница 37: ...as been damaged by accident or misuse exposed to radioactive or corrosive materials has foreign material penetrating to the inside of the product or as a result of service or modification by other tha...

Страница 38: ......

Страница 39: ......

Страница 40: ...haus Corporation 7 Campus Drive Suite 310 Parsippany NJ 07054 USA Tel 1 973 377 9000 Fax 1 973 944 7177 www ohaus com P N 30529806 A 2019 Ohaus Corporation all rights reserved Printed in China 3052980...

Отзывы: