EN-14
3
000 Series Indicators
3.3.2 Span Calibration
Span Calibration uses one point. The span point is established with a calibration mass placed on the scale.
When SPAN is displayed, press the
Yes
button to access the Span Calibration menu item.
The display flashes the span calibration point. Place the specified weight on the scale and press the
Yes
button.
To choose a different span point, repeatedly press the
No
button to increment the selections or press the
Back
button to decrement the selections. Refer to Table 3-3 for available span points. When the desired value is
displayed, place the specified weight on the scale and press the
Yes
button.
The display shows --C-- while the span point is established.
If span calibration was successful, the scale exits to the next Calibration menu and displays [LINEAr]
Note: Span Calibration should be performed after Zero Calibration.
3.3.3 Linearity Calibration
Linearity calibration uses 3 calibration points. The first calibration point is established with no weight on the
scale. The second calibration point is established at approximately half capacity. The third calibration point
is established at capacity. The Linearity calibration points are fixed and cannot be altered by the user during
the calibration procedure. Refer to Table 3-3 for the linearity points.
When LINEAr is displayed, press the
Yes
button to access the Linearity Calibration menu item.
The display flashes 0. With no weight on the scale, press the
Yes
button to establish the zero point.
The display shows --C-- while the zero point is established.
The display flashes the mid calibration point.
Place the specified weight on the scale and press the
Yes
button.
The display shows --C-- while the mid point is established.
The display flashes the full calibration point.
Place the specified weight on the scale and press the
Yes
button.
The display shows --C-- while the full point is established.
If linearity calibration was successful, the scale exits to the next Calibration menu and displays [GEO]