EN-2                         ORP electrodes                                     

 

Performing ORP measurement: 

1. After removing the electrode from the 

storage solution bottles, rinse the ORP 

electrode with pure water before 

measurement; dip the ORP electrode into the 

sample.  
2. Connect the ORP electrode to the pH 

meter, set the meter at mV measurement 

mode. 
3. If the ORP electrode is refillable (STORP2), 

slide down the red cover to expose the filling 
hole when measurement, remember to close 

it after measurement to avoid leakage. 
4. After measurement, rinse the ORP 

electrode with pure water and store back in 

the small protection bottle with soaking 
solution. 

ORP electrode quality check: 

To check if the ORP electrode quality is good: 
1. Prepare 2 standard buffers solutions for 

ORP: Saturated quinhydrone pH4.01 and 
pH7.00. 
2. Dip the ORP electrode into buffer 7.00, the 

mV value should be 

86mV ± 15mV

3. After rinsing in pure water, dip the ORP 

electrode into buffer 4.01, the difference 

between 7.00 and 4.01 should be 

≥ 165mV.

 

4. If the check result is not good, should 

clean the ORP electrode then check again; or 

purchase a new one.

 

 

 

 

Содержание STORP1

Страница 1: ...ustin Ft Worth NicolScales com 800 225 8181 Contact Us Nicol Scales Measurement is an ISO Accredited Calibration Company that has provided calibration repair and sales of all types of weighing and mea...

Страница 2: ...duction PRO Ohaus Corporation 7 Campus Drive Suite 310 Parsippany NJ 07054 USA Tel 973 377 9000 Fax 973 944 7177 With offices worldwide Con oficinas en todo el mundo Avec des bureaux dans le monde ent...

Страница 3: ...EN 2 ORP electrodes...

Страница 4: ...ence Electrolyte Gel 3M KCl STORP1 Solution 3M KCl STORP2 Zero potential value 86mV 15mV Grade difference 165mV The ORP electrode is also called Redox electrode measures the Oxidation Reduction Potent...

Страница 5: ...after measurement to avoid leakage 4 After measurement rinse the ORP electrode with pure water and store back in the small protection bottle with soaking solution ORP electrode quality check To check...

Страница 6: ...should clean the electrode as follows 1 Inorganic contamination dip the ORP electrode into 0 1M HCl for more than 10 minutes then rinse with pure water Soak the ORP electrode in 3M KCl solution for m...

Страница 7: ......

Страница 8: ...rio STORP2 Rango de temperatura 0 80 STORP1 0 100 STORP2 Electr lito de referencia Gel 3M KCl STORP1 Soluci n 3M KCl STORP2 Valor potencial cero 86 mV 15 mV Diferencia de grado 165 mV El electrodo ORP...

Страница 9: ...el orificio de llenado durante la medici n y recuerde cerrarlo despu s de la medici n para evitar goteo 4 Despu s de la medici n enjuague el electrodo POR con agua pura y vuelva a almacenarlo en el fr...

Страница 10: ...el electrodo de la siguiente manera 1 Contaminaci n inorg nica Sumerja el electrodo POR en HCl 0 1M durante m s de 10 minutos y luego enjuague con agua pura Enjuague el electrodo POR en soluci n 3M KC...

Страница 11: ......

Страница 12: ...f rence Gel 3M KCl STORP1 Solution 3M KCl STORP2 Valeur potentielle du z ro 86 mV 15 mV Diff rence selon la qualit 165mV L lectrode de Potentiel d Oxydo R duction porte aussi le nom de Redox elle mesu...

Страница 13: ...refermer apr s la mesure afin d viter des fuites 4 Apr s la mesure rincer l lectrode PRO l eau pure et la ranger dans la petite bouteille de protection avec une solution de stockage V rification de l...

Страница 14: ...ante 1 Contamination inorganique tremper l lectrode PRO dans de l HCl 0 1 M pendant plus de 10 minutes et la rincer ensuite dans de l eau pure Faire tremper l lectrode de potentiel oxydo r ducteur dan...

Отзывы: