OHAUS STORP1 Скачать руководство пользователя страница 4

ORP electrodes                         EN-1

 

Model 

Item No. 

Description 

STORP1  30038555 

Not refillable, plastic 

shaft with Platinum 

Disk 

STORP2  30038553 

Refillable, glass 

shaft with Platinum 

Band 

3M KCl 

30059256 

Protection solution 

 

Specifications: 

 

Connection:        

BNC 

Cable length:   

1.0m 

Shaft length:    

120mm 

Shaft diameter:   

12mm 

Shaft material:   

 

 

Plastic  (STORP1) 

 

Glass  (STORP2) 

Temperature range:   

 

0-80

   (STORP1) 

 

0-100

  (STORP2) 

Reference Electrolyte: 

 

Gel 3M KCl (STORP1) 

 

Solution 3M KCl (STORP2) 

Zero potential value: 

 

      

86mV± 15mV 

Grade difference: 

≥ 165mV 

 

 

The ORP electrode is also called Redox 

electrode, measures the   

O

xidation-

R

eduction 

P

otential, related to the overall 

availability of electrons in the media, 

specifically the ratio of positive and negative 

ions in the solution. 

 

Normal application areas are in water quality 

testing for swimming pools, fish farming and 

aquariums etc. 
ORP is expressed in millivolts (mV). A range 
of -1000 mV to 1000mV is common with most 

ORP test.   
The pH value influences the ORP value 

significantly.

 

Содержание STORP1

Страница 1: ...ustin Ft Worth NicolScales com 800 225 8181 Contact Us Nicol Scales Measurement is an ISO Accredited Calibration Company that has provided calibration repair and sales of all types of weighing and mea...

Страница 2: ...duction PRO Ohaus Corporation 7 Campus Drive Suite 310 Parsippany NJ 07054 USA Tel 973 377 9000 Fax 973 944 7177 With offices worldwide Con oficinas en todo el mundo Avec des bureaux dans le monde ent...

Страница 3: ...EN 2 ORP electrodes...

Страница 4: ...ence Electrolyte Gel 3M KCl STORP1 Solution 3M KCl STORP2 Zero potential value 86mV 15mV Grade difference 165mV The ORP electrode is also called Redox electrode measures the Oxidation Reduction Potent...

Страница 5: ...after measurement to avoid leakage 4 After measurement rinse the ORP electrode with pure water and store back in the small protection bottle with soaking solution ORP electrode quality check To check...

Страница 6: ...should clean the electrode as follows 1 Inorganic contamination dip the ORP electrode into 0 1M HCl for more than 10 minutes then rinse with pure water Soak the ORP electrode in 3M KCl solution for m...

Страница 7: ......

Страница 8: ...rio STORP2 Rango de temperatura 0 80 STORP1 0 100 STORP2 Electr lito de referencia Gel 3M KCl STORP1 Soluci n 3M KCl STORP2 Valor potencial cero 86 mV 15 mV Diferencia de grado 165 mV El electrodo ORP...

Страница 9: ...el orificio de llenado durante la medici n y recuerde cerrarlo despu s de la medici n para evitar goteo 4 Despu s de la medici n enjuague el electrodo POR con agua pura y vuelva a almacenarlo en el fr...

Страница 10: ...el electrodo de la siguiente manera 1 Contaminaci n inorg nica Sumerja el electrodo POR en HCl 0 1M durante m s de 10 minutos y luego enjuague con agua pura Enjuague el electrodo POR en soluci n 3M KC...

Страница 11: ......

Страница 12: ...f rence Gel 3M KCl STORP1 Solution 3M KCl STORP2 Valeur potentielle du z ro 86 mV 15 mV Diff rence selon la qualit 165mV L lectrode de Potentiel d Oxydo R duction porte aussi le nom de Redox elle mesu...

Страница 13: ...refermer apr s la mesure afin d viter des fuites 4 Apr s la mesure rincer l lectrode PRO l eau pure et la ranger dans la petite bouteille de protection avec une solution de stockage V rification de l...

Страница 14: ...ante 1 Contamination inorganique tremper l lectrode PRO dans de l HCl 0 1 M pendant plus de 10 minutes et la rincer ensuite dans de l eau pure Faire tremper l lectrode de potentiel oxydo r ducteur dan...

Отзывы: