EN-14 STARTER 5000 Bench pH Meter
5
SETUP
5.1
System Setting
System setting is used to set meter parameters, the meter information could also be found in it.
Please see section 3.4 for reference.
5.1.1
System setting
►Save Mode
Save mode has 2 options, one is Auto Save Endpoint Reading, another is Manual Save Endpoint
Reading. Auto Save Endpoint Reading means that when the measurement reaches the endpoint
(auto, manual or time endpoint), the meter will save the endpoint reading automatically.
The Save icon
will change from black to light grey
; An endpoint reading can be
saved only once.
5.1.2
System Setting
►Output
When selecting shut off connection, the main interface will display the save icon
, or
display the Output Icon
.
5.1.3
System Setting
►
Restore Factory Setup
Restoring the factory setup includes the following:
1.
Shut off GLP mode
2.
Shut off Auto Save mode
3.
Shut off continuous measure
4.
Shut off (output) connection
5.
Shut off calibration Reminder
6.
The resolution is: 0.001pH, 0.01mV
7.
Measurement mode: pH
8.
Self buffer group will be deleted
9.
Baud rate: 9600
10. Brightness of screen: 70%
11. Shut off Screen Protection
12. MTC: 25°C
13. Endpoint: Auto
14. Time endpoint: 10 Second
15. Delete all the calibration data
5.1.4
System Setting
►
Service Password
Not used.