CHAPTER 2 DIAGNOSIS
Ohaus Corporation
www.ohaus.com
2-4
ST5000 Bench Top pH Meter
Service Manual
2.7 DIAPHRAGM CONTAMINATION ISSUE.
Most of the problems during pH measurements are caused through a contaminated diaphragm,
Symptoms:
Dependancy of the stirrer speed
Drifting result
Shift of offset
Decrease of slope
Contaminated diaphragm = Number one troublemaker!
Diaphragm contamination can occur with:
solutions that contaminate the diaphragm
(e.g. solutions containing proteins, paints, AgS,…)
suspensions, emulsions
highly acidic solutions
strong alkaline solutions
low ionic samples (distilled water, ultrapure water)
solutions with high salt concentration
non-aqueous solutions
Cleaning the diaphragm
blocked with
AgCl
-
with concentrated ammonia
blocked with
Ag
2
S
-
with 8 % thiourea in 0.1 molar HCl
blocked with
proteins
-
with 5 % pepsin in 0.1 molar HCl
other contaminations
-
in ultrasonic bath with water or 0.1 molar HCl
Or just replace with a new one
2.8 RECONDITIONING OF THE DRY GLASS MEMBRANE.
Dry membrane:
Re-conditioning in 0.1 mol/L HCl during 12 hours.
2.9 LIFETIME OF AN ELECTRODE.
The lifetime of an electrode is limited. Apart from maintenance and sample consistency,
one of the reasons is the aging of the membrane.
The aging o the pH sensitive membrane and therefore the decrease of the H
+
-sensitivity
depends on the temperautre: