CHAPTER 2 DIAGNOSIS
Ohaus Corporation
www.ohaus.com
2-2
ST5000 Bench Top pH Meter
Service Manual
TABLE 2-2. EQUIPMENT TOUCH PANEL DOES NOT RESPONCE.
PROBABLE CAUSE
REMEDY
Touch Panel controls
does not respond
Bad touch panel calibration
Lose cable connection.
Re-calibrate Touch Panel (A.4)
Check film cable connection from main
PCBA to LCD screen.
Check film cable connection from touch
panel to LCD screen.
TABLE 2-3 MEASURING RESULTS INCONSISTANT.
PROBABLE CAUSE
REMEDY
End results are not
accurate.
Inconsistent pH reading.
Inconsistent
temperature
reading.
Wrong buffer group is used
during calibration
Recalibrate the pH electrode and
temperature electrode.
Use the correct buffer group during
calibration.