a temperature probe is used,
ATC
and the sample temperature are displayed. If the meter does
not detect a temperature probe, it automatically switches to the manual temperature
compensation mode and MTC appears. MTC temperature should be set. (See 4.2, 4.3)
Note:
The meter accepts
NTC 30 kΩ temperature sensor.
3.4
Using the memory
3.4.1
Storing a reading
The STARTER 300 pH meter can store up to 30 endpointed results. Press
Store
when the
measurement has endpointed.
M01
indicates that one result has been stored.
If you press
Store
when
M30
is displayed,
FUL
indicates the memory is full. To store further data
you will have to clear the memory (see below).
3.4.2
Recalling from memory
Press and hold
Recall
to recall the stored values from memory when the current measurement
has endpointed.
Press
up
or
down
buttons to scroll through the stored results.
R01
to
R30
indicates which result is
being displayed. Press
Read
to exit.
3.4.3
Clearing the memory
Continue pressing
up
or
down
buttons to scroll through the stored results until
MRCL
appears.
Then press
Read
,
CLr
blinks. Press
Read
again to confirm the deletion or press
Exit
to return to
the measurement mode without deleting the data.
4
Setup
4.1
Set endpoint mode
The automatic endpoint
is the default setting of the meter. By pressing and holding
Read
,
you can switch between the auto and manual endpoint modes. In auto endpoint mode, press
Read
to start the measurement, then the meter judge the endpoint automatically and lock the
endpoint reading.
In manually endpoint mode, you should press
Read
one more time to confirm the final reading
result: the sample reading freezes, the measurement icon blinks 3 times and disappears. Then,
the endpoint stability icon
blinks 3 times and freezes on the display.
Stability criterion for pH and mV measurement
– The signal of the sensor input may not
change by more than 0.1mV in 6 seconds.
4.2
Set temperature unit
If the meter does not detect a temperature probe, it automatically switches to the manual
temperature compensation mode and
MTC
appears.
To set the
MTC
temperature and the temperature unit, press and hold
Mode/Setup
until the setup
icon
appears on the display and the current temperature unit blinks. Use
up
or
down
buttons to
switch between
°
C and
°
F . Press
Read
to confirm your selection. Continue with MTC
temperature setting or press
Exit
to go back to the measurement screen.