OHAUS STARTER 300 Скачать руководство пользователя страница 10

 

3.1.3 

Performing 2-point calibration 

• Step 1: Perform the first calibration as described above in "

Performing 1-point calibration

". 

• Step 2: Rinse the electrode with pure water. 
• Step 3: Place the electrode in the next calibration buffer and press 

Cal

 

 

The  calibration  icon 

 and  the  measurement  icon   appear  on  the  display.  The  measurement 

icon is blinking during calibration measurement. The meter endpoints according to the preselected 
endpoint mode after the signal has stabilized or after pressing 

Read

.  

 
When  the  calibration  point  is  finished,  the  relevant  buffer  value  is  displayed  and  stored;  the 
measurement icon blinks 3 times and disappears. Then, the endpoint stability and  auto endpoint 
icons 

 blink  3  times  and  freeze  on  the  display  and 

Cal  2

”  or  “

Cal  3

”  stops  blinking.  If  the 

calibration  was  endpointed  manually,  then  only  endpoint  stability  icon 

 blinks  3  times  and 

freezes on the display. 
 
To finish the calibration and return to the sample measurement, press 

Read

. The offset value and 

the slope are then shown on the display for 3 seconds. To reject the calibration, press 

Exit

 before 

the meter returns automatically after 3 seconds to the measurement screen. 
 

Note: 

The  use  of  a  temperature  sensor  or  an  electrode  with  a  built-in  temperature  sensor  is 
recommended.  If  you  use  the  MTC  mode,  you  should  enter  the  correct  temperature  value  and 
keep all buffer and sample solutions at the set temperature.  (see 

4.2

To ensure the most accurate pH measurement, you should perform a calibration once a day. 

3.2 

Sample measurement 

3.2.1 

pH measurement 

Place  the  electrode  in  the sample  and  press 

Read

 to start the measurement.  The measurement 

icon   appears  on  the  display.  The  measurement  icon  is  blinking  during  measurement  and  the 
display shows the pH value of the sample. The automatic endpoint 

 is the default setting of 

the  meter.  When  the  signal  has  stabilized,  the  sample  reading  freezes,  the  measurement  icon 
blinks 3 times and disappears. Then, the endpoint stability and auto endpoint icons 

 blink 3 

times and freeze on the display.  
 
By pressing and holding 

Read

, you can switch between the auto and manual endpoint modes. To 

manually  endpoint  a  measurement,  press 

Read

:  the  sample  reading  freezes,  the  measurement 

icon  blinks  3  times  and  disappears.  Then,  the  endpoint  stability  icon 

 blinks  3  times  and 

freezes on the display. 
 

Stability  criterion  for  pH  and  mV  measurement

 

–  The  signal  of  the  sensor  input  may  not 

change by more than 0.1mV in 6 seconds. 

3.2.2 

mV measurement 

To switch to mV mode, press 

Mode

To perform a mV measurement, follow the same procedure as for pH measurement. 
 

3.3 

Temperature measurement 

For better accuracy, we recommend the use of either a built-in or a separate temperature probe. If 

Содержание STARTER 300

Страница 1: ...STARTER 300 Portable pH Meter Instruction Manual...

Страница 2: ......

Страница 3: ...libration 7 3 1 3 Performing 2 point calibration 8 3 2 Sample measurement 8 3 2 1 pH measurement 8 3 2 2 mV measurement 8 3 3 Temperature measurement 8 3 4 Using the memory 9 3 4 1 Storing a reading 9...

Страница 4: ...permanent installations Integrated labeling design for quick product identification IP54 protection against dust and water Ohaus renowned user friendly software 1 1 Safety measures Measures for your...

Страница 5: ...endpoint 3 Measurement icon measurement or calibration is running 4 Calibration icon calibration in progress 5 Setup icon instrument is in the setup mode 6 Battery status icon shows the condition of t...

Страница 6: ...Start calibration Review the latest calibration data Meter on Back to measurement screen Meter off Store current reading to memory Increase value during setting Scroll up through the memory Recall sto...

Страница 7: ...the batteries 1 Remove the battery cover screw and remove the battery cover 2 Insert the batteries in the battery compartment as shown 3 Replace the battery cover and tighten the battery cover screw...

Страница 8: ...p forward to lock in position Slide the shaft of the sensor into the clip from above 2 4 Installing the IP54 cover Use the small plastic tool as shown 2 5 Integrated stand for table top use One unique...

Страница 9: ...libration buffer and press Cal Cal 1 is blinking during calibration The calibration icon and the measurement icon appear on the display The measurement icon is blinking during calibration measurement...

Страница 10: ...is recommended If you use the MTC mode you should enter the correct temperature value and keep all buffer and sample solutions at the set temperature see 4 2 To ensure the most accurate pH measuremen...

Страница 11: ...confirm the deletion or press Exit to return to the measurement mode without deleting the data 4 Setup 4 1 Set endpoint mode The automatic endpoint is the default setting of the meter By pressing and...

Страница 12: ...t out of range Make sure the buffer is correct and fresh Clean or replace the pH electrode Error 5 Slope out of range Make sure the buffer is correct and fresh Clean or replace the pH electrode Error...

Страница 13: ...e to press the other keys until all the icons have disappeared When the self diagnosis has been completed successfully PAS appears If self diagnosis fails error message Err 1 appears Note You have to...

Страница 14: ...9 03 10 28 50 1 98 4 06 6 97 8 99 10 10 Buffer group b3 JJG119 Temp pH1 68 pH4 00 pH6 86 pH9 18 pH12 46 5 1 67 4 00 6 95 9 39 13 21 10 1 67 4 00 6 92 9 33 13 01 15 1 67 4 00 6 90 9 28 12 82 20 1 68 4...

Страница 15: ...Corporation 7 Campus Drive Suite 310 Parsippany NJ 07054 USA Tel 973 377 9000 Fax 973 944 7177 With offices worldwide www ohaus com PN 83033978 Ohaus Corporation 2011 all rights reserved 83033978 121...