OHAUS STARTER 2100 Скачать руководство пользователя страница 9

 

Installation

 

Carefully unpack the box.  

2.1 

Package contents 

The model ST2100-B (basic model) has the following items:

  

 
 
 
 
 
 

In addition to ST2100-B content, the model ST2100-E package also includes the following: 

 
 
 
 

 In addition to ST2100-E content, the model ST2100-F package also includes the following: 

 
 
 

You also could order other Ohaus pH electrodes for different applications. 
Each pH buffer powder should be dissolved in 250ml pure water or deionized water in volumetric 
flask. You also can order the bottle buffer solution from Ohaus. 
 
Accessories: 

Model 

Description 

Item NO 

ST310 

3-in-1 plastic refillable pH Electrode  

83033965 

ST210 

2-in-1 plastic refillable pH Electrode  

83033966 

ST320 

3-in-1 plastic gel pH Electrode(no need to refill) 

83033967 

ST230 

2-in 1 glass muddy sample pH Electrode  

83033968 

STPURE 

2-in-1 glass pure water sample pH Electrode  

83033969 

STTEMP30 

Temperature Electrode  

83033970 

STORP1 

Gel plastic ORP electrode  

30038555 

STORP2 

Refillable glass ORP electrode  

30038553 

STREF01 

Ag/AgCl reference electrode  

30059253 

 

Buffer powder set (4.01; 7.00; 10.0) 

83033971 

Buffer pH4.01  250ml 

30065083 

Buffer pH7.00  250ml 

30065084 

Buffer pH10.00  250ml 

30065086 

pH electrode Reference Electrolyte 

30059255 

ST2100-B 

Units 

STARTER 2100

  

Built-in electrode arm 

12V Power supply 

1 set 

ST210 2-in-1 pH electrode 

pH Buffer Power Set (4.01, 7.00, 10.01) 

1 set 

STTEMP30  temperature electrode 

Содержание STARTER 2100

Страница 1: ...STARTER 2100 Bench pH Meter Instruction Manual...

Страница 2: ......

Страница 3: ...tion 8 3 1 Calibration 8 3 1 1 Buffer group 8 3 1 2 Performing 1 point calibration 9 3 1 3 Performing 2 point calibration 9 3 2 Sample measurement 10 3 2 1 pH measurement 10 3 2 2 mV measurement 10 3...

Страница 4: ......

Страница 5: ...show the pH electrode performance Can recall last calibration data Quick Guide attached in the meter to help user operate 1 1 Safety measures Measures for your protection Never work in an environment...

Страница 6: ...when the equipment is operated in a commercial environment This equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance with the instruction manual...

Страница 7: ...trode condition is not good or needs cleaning 2 Measurement icon measurement or calibration is running disappear means the reading is locked no matter measurement 3 Calibration icon 1 point or 2 point...

Страница 8: ...ting Start calibration Review the latest calibration data slope offset Meter turn on Exit return to measurement screen Meter turn off Entered temperature setup mode Increase temp value when in temp se...

Страница 9: ...able pH Electrode 83033965 ST210 2 in 1 plastic refillable pH Electrode 83033966 ST320 3 in 1 plastic gel pH Electrode no need to refill 83033967 ST230 2 in 1 glass muddy sample pH Electrode 83033968...

Страница 10: ...ectrode arm on the left or right sight of the STARTER 2100 meter Remove the arm hole rubber cover from the meter Insert the metal stick into the hole and screw to fasten it Install the upper electrode...

Страница 11: ...pH electrode The BNC pH Cinch Temp and banana REF socket For ST210 we only need to connect the pH electrode with the pH socket BNC For STTEMP30 connect with Temp Socket Cinch For ST310 connect both t...

Страница 12: ...le or buffer solution user should stir several seconds then wait 30 to 60 seconds for the signal to be stable and then press the button to operate do Calibration or measurement 3 1 Calibration 3 1 1 B...

Страница 13: ...libration return to the measurement screen Press button Cal to do the 2 point calibration Note With the 1 point calibration only the offset is adjusted If the pH electrode was previously calibrated wi...

Страница 14: ...r accuracy we recommend to use temperature electrode If a temperature electrode is used ATC Auto Temperature Compensation and the sample temperature are displayed If the meter does not detect a temper...

Страница 15: ...some organic solvents such as toluene xylene and methyl ethyl ketone MEK Any spillage should be immediately wiped off 5 3 Electrode maintenance Make sure the electrode is filled with electrolyte solut...

Страница 16: ...message Err appears Note You have to finish pressing all five keys within 2 minutes otherwise Err appears and you will have to repeat the procedure 5 5 Recover factory settings When the meter is off...

Страница 17: ...5 C Calibration 1 or 2 points 1 predefined buffer group 4 01 7 00 10 01 Memory Recall last calibration data Power supply 110 240V 50Hz DC 12V Size weight Approximately 220 W x 175 D x 78 H mm 0 75 kg...

Страница 18: ...10 10 Buffer group JJG119 Temp pH1 68 pH4 00 pH6 86 pH9 18 pH12 46 5 1 67 4 00 6 95 9 39 13 21 10 1 67 4 00 6 92 9 33 13 01 15 1 67 4 00 6 90 9 28 12 82 20 1 68 4 00 6 88 9 23 12 64 25 1 68 4 00 6 86...

Страница 19: ...rporation 7 Campus Drive Suite 310 Parsippany NJ 07054 USA Tel 973 377 9000 Fax 973 944 7177 With offices worldwide www ohaus com pH ohaus com PN 30031622 A Ohaus Corporation 2013 all rights reserved...

Отзывы: