OHAUS STARTER 2100 Скачать руководство пользователя страница 13

 

3.1.2 

Performing 1-point calibration 

Calibration:

 pH electrode need to be calibrated with pH standard buffer solution before a proper 

pH measurement.

  Calibration

 is to display the right 

pH

 value when meter receive the 

mV 

value 

signal from the pH electrode.  

Slope: the linear coefficient between mV and pH according to theoretical value (e.g.   -
59.16mV/pH @ 25°C which means 100% slope); 
Offset: the mV value when pH value is 7.00. (Theoretical value is 0 mV); 

 
When 

STARTER  2100

  is  in 

pH  measurement  mode

  (see  3.2),  we  can  do  the  pH  electrode 

calibration; place the pH electrode in a calibration buffer: 

 

Press button-

Cal

Cal 

1

” displays on the top right of the screen and is blinking. Measurement 

icon   appear on the top of the screen,    is blinking during calibration.  

 

If the reading is stable,  press button-

Read

 to lock the reading and finish 1 point calibration. 

The  buffer  value  (e.g.  pH  7.00)  with  the  temperature  display  on  the  screen,  the  meter  can 
recognize the buffer automatically (auto buffer recognition). 
 

The 1-point calibration is finished; now we have 3 choices to do next: 

 

Press button-

Read

 to store the 1-point calibration data and exit, the 

offset

 and the 

slope

 are 

shown on the display for 3 seconds then return to the measurement screen. 

 

Press button-

Exit

 to reject the calibration, return to the measurement screen. 

 

Press button-

Cal

 to do the 2-point calibration. 

 

Note: 

With the  1-point calibration  only the 

offset 

is  adjusted.  If the  pH  electrode  was  previously 

calibrated  with  multi-point  calibration  the  previously  stored 

slope

  will  remain.  Otherwise 

theoretical 

100% slope

 (-59.16 mV / pH @25

°

C) will be used. 

3.1.3 

Performing 2-point calibration 

 

Perform 1-point calibration as described above. 

 

Rinse the pH electrode with pure water and wipe off water with tissue. 

 

Place the electrode in the next calibration buffer and press button-

Cal

,

 

Cal 

2

” displays on the 

left bottom of the screen and 

Cal 

” is blinking. 

 

and   appear on the top of the screen,    

is blinking during calibration. 

 

When reading is stable, press button-

Read

 to lock the reading and finish 2-point calibration, 

 disappear. The buffer pH value (e.g.  

7.00 pH

) with the temperature display on the screen. 

 

The 2-point calibration is finished, and then we have 2 choices to do next: 

 

Press  button-

Read

  to  store  the  2-point  calibration,  the 

offset

  and 

slope

  are  shown  on  the 

display for 3 seconds then return to the measurement screen. 

 

Press button-

Exit

 to reject the calibration, return to the measurement screen. 

 

Note: 

The  use  of  a  temperature  electrode  (e.g.  STTEMP30)  or  a  pH  electrode  with  a  built-in 

temperature sensor (3-in-1 pH electrode) is recommended. If you use the 

MTC

 mode, you should 

enter  the  correct  temperature  value  and  keep  all  buffer  and  sample  solutions  at  the  set 
temperature. (See 4.1) 

Содержание STARTER 2100

Страница 1: ...STARTER 2100 Bench pH Meter Instruction Manual...

Страница 2: ......

Страница 3: ...tion 8 3 1 Calibration 8 3 1 1 Buffer group 8 3 1 2 Performing 1 point calibration 9 3 1 3 Performing 2 point calibration 9 3 2 Sample measurement 10 3 2 1 pH measurement 10 3 2 2 mV measurement 10 3...

Страница 4: ......

Страница 5: ...show the pH electrode performance Can recall last calibration data Quick Guide attached in the meter to help user operate 1 1 Safety measures Measures for your protection Never work in an environment...

Страница 6: ...when the equipment is operated in a commercial environment This equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance with the instruction manual...

Страница 7: ...trode condition is not good or needs cleaning 2 Measurement icon measurement or calibration is running disappear means the reading is locked no matter measurement 3 Calibration icon 1 point or 2 point...

Страница 8: ...ting Start calibration Review the latest calibration data slope offset Meter turn on Exit return to measurement screen Meter turn off Entered temperature setup mode Increase temp value when in temp se...

Страница 9: ...able pH Electrode 83033965 ST210 2 in 1 plastic refillable pH Electrode 83033966 ST320 3 in 1 plastic gel pH Electrode no need to refill 83033967 ST230 2 in 1 glass muddy sample pH Electrode 83033968...

Страница 10: ...ectrode arm on the left or right sight of the STARTER 2100 meter Remove the arm hole rubber cover from the meter Insert the metal stick into the hole and screw to fasten it Install the upper electrode...

Страница 11: ...pH electrode The BNC pH Cinch Temp and banana REF socket For ST210 we only need to connect the pH electrode with the pH socket BNC For STTEMP30 connect with Temp Socket Cinch For ST310 connect both t...

Страница 12: ...le or buffer solution user should stir several seconds then wait 30 to 60 seconds for the signal to be stable and then press the button to operate do Calibration or measurement 3 1 Calibration 3 1 1 B...

Страница 13: ...libration return to the measurement screen Press button Cal to do the 2 point calibration Note With the 1 point calibration only the offset is adjusted If the pH electrode was previously calibrated wi...

Страница 14: ...r accuracy we recommend to use temperature electrode If a temperature electrode is used ATC Auto Temperature Compensation and the sample temperature are displayed If the meter does not detect a temper...

Страница 15: ...some organic solvents such as toluene xylene and methyl ethyl ketone MEK Any spillage should be immediately wiped off 5 3 Electrode maintenance Make sure the electrode is filled with electrolyte solut...

Страница 16: ...message Err appears Note You have to finish pressing all five keys within 2 minutes otherwise Err appears and you will have to repeat the procedure 5 5 Recover factory settings When the meter is off...

Страница 17: ...5 C Calibration 1 or 2 points 1 predefined buffer group 4 01 7 00 10 01 Memory Recall last calibration data Power supply 110 240V 50Hz DC 12V Size weight Approximately 220 W x 175 D x 78 H mm 0 75 kg...

Страница 18: ...10 10 Buffer group JJG119 Temp pH1 68 pH4 00 pH6 86 pH9 18 pH12 46 5 1 67 4 00 6 95 9 39 13 21 10 1 67 4 00 6 92 9 33 13 01 15 1 67 4 00 6 90 9 28 12 82 20 1 68 4 00 6 88 9 23 12 64 25 1 68 4 00 6 86...

Страница 19: ...rporation 7 Campus Drive Suite 310 Parsippany NJ 07054 USA Tel 973 377 9000 Fax 973 944 7177 With offices worldwide www ohaus com pH ohaus com PN 30031622 A Ohaus Corporation 2013 all rights reserved...

Отзывы: