B-1

A

PPENDIX B FACTORY CALIBRATIONS

B.FACTORY  LINEARITY  CALIBRATIONS

The Scout Pro service and factory linearity calibrations use three calibration points; zero, mid-scale

and full scale.

B.1 SERVICE  CALIBRATION

This  calibration  is  used  when  an  error  2  has  been  displayed  by  the  balance.  The  calibration  is

performed with no error checking and is saved as a user calibration. It is not a factory calibration

because no data is written to the EEPROM.

1.

Start with the unit off.

2.

Press and hold both the 

ON/ZERO 

Off

 and the 

PRINT 

Unit

  buttons until the display reads

rAMP.

3.

Press and release the 

PRINT 

Unit

 button. The display will show 

Lin

.

4.

Press and release the 

On/Zero 

Off

 button. The display will show -

C-

 blinking and then the

first mass value.

5.

Place masses totaling this value on the balance and press the 

On/Zero 

Off

 

 button.

6.

Press and release the 

On/Zero 

Off

  button. The display will show -

C-

 blinking and then the

second mass value.

7.

Place masses totaling this value on the balance and press the 

On/Zero 

Off

 

 button.

8.

Press and release the 

On/Zero 

Off

 

 button. The display will show -

C-

 blinking and then

donE

 followed by 

Lin

.

9.

Repeatedly press and release the 

PRINT 

Unit

 

button until the display shows 

END

.

10. Press and release the 

On/Zero 

Off

  button. The balance will return to the weighing mode.

B.2 FACTORY CALIBRATION

This calibration must be used when either a PC board or a transducer is replaced.

1.

Remove power from the balance.

2.

Install an RS232 option with the pins shorted. See Appendix A.3.  If the RS232 option is

not available, pins 2 and 7 of IC9 may be shorted using a micro alligator clip.  See Figure

B-1.

CAUTIONS

1. Be very careful that no other pins are shorted, this could destroy the IC.

2. Do not attempt to solder the jumper.

www.mirvesov.ru

Тел

.: (495) 921-44-57

All manuals and user guides at all-guides.com

Содержание Scout pro

Страница 1: ...SERVICE MANUAL SCOUT PRO BALANCES Ohaus Corporation 19A Chapin Road P O Box 2033 Pine Brook NJ 07058 2033 973 377 9000 All manuals and user guides at all guides com a l l g u i d e s c o m...

Страница 2: ...www mirvesov ru 495 921 44 57 All manuals and user guides at all guides com...

Страница 3: ...ration assumes no liability arising from the use or misuse of this material Reproduction of this material is strictly prohibited Material in this manual is subject to change Copyright 2003 Ohaus Corpo...

Страница 4: ...www mirvesov ru 495 921 44 57 All manuals and user guides at all guides com...

Страница 5: ...nt 3 2 3 1 3 Transducer Replacement 3 4 CHAPTER 4 TESTING 4 1 Testing 4 1 4 1 1 Operation Test 4 1 4 1 2 Segment Display Test 4 1 4 2 Performance Tests 4 1 4 2 1 Repeatability Test 4 2 4 2 2 Off Cente...

Страница 6: ...t Board removal 3 2 3 4 Transducer Removal 3 4 4 1 LCD Full Display 4 1 4 2 Transducer Front View Down Stop Adjustment 4 4 5 1 Scout Pro Exploded View 5 2 A 1 Screw and Housing Removal A 2 A 2 Connect...

Страница 7: ...uld be familiar with the Instruction Manual which is packed with everybalance The procedures inthis manual assumes the technicianperforming themhas a working knowledge of the use of standard hand tool...

Страница 8: ...d as follows 1 3 1 Special Tools and Test Equipment List 1 Alternate voltage Power Adapter if local power requirements do not match Balance Adapter voltage ratings 2 Scout Pro RS232 option PN 71147376...

Страница 9: ...00 400 400 600 2000 4000 6000 Readability g 0 01 0 1 1 0 Repeatability Std Dev g 0 01 0 1 1 0 Off Center Loading g 0 01 0 1 1 0 Linearity g 0 01 0 1 1 0 Tare range To capacity by subtraction Over Rang...

Страница 10: ...1 4 CHAPTER 1 INTRODUCTION www mirvesov ru 495 921 44 57 All manuals and user guides at all guides com...

Страница 11: ...omponents and operation Do not attempt repairs unless you fully understand the operation of the balance 2 2 1 Diagnosis 1 Isolate and identify the symptom 2 Refer to Table 2 1 Diagnostic guide and loc...

Страница 12: ...STIC GUIDE PROBABLE CAUSE 1 No power 2 Power Jack defective 3 Weak battery 4 Battery connector poor contact 5 PC Board defective 1 Weak battery 2 AC Adapter defective or wrong type being used 3 PC Boa...

Страница 13: ...isaligned 1 Balance not warmed up 2 Defective Transducer 3 PC Board defective 1 Balance has detected an error condition 2 Defective Transducer 3 PC Board defective REMEDY 4 Replace Transducer 5 Replac...

Страница 14: ...ct mass Err 5 Parts counting error 1 Sample is less than 1d Change sample quantity or increase sample Err 8 RS232 buffer is full 1 Check printing device Err 9 Internal data error 1 Restore data file 2...

Страница 15: ...ponents to static electricity discharge When servicing a wrist Ground Strap with a 10 Megohm series resistor to earth ground is highly recommended The Printed Circuit Board 9 should be handled by gras...

Страница 16: ...t should be replaced rather than repaired To replace the Printed Circuit Board 9 proceed as follows 1 Remove the balance cover refer to paragraph 3 1 1 CAUTION WHEN HANDLING THE PRINTED CIRCUIT BOARD...

Страница 17: ...sure that the Printed Circuit Board is positioned under the 2 clips on the Base 10 9 Connect the flexible cable from the Transducer to Connector J6 on the Printed Circuit Board 9 10 Connect the RS232...

Страница 18: ...can render it inoperative 1 Remove the cover in accordance with the procedures in paragraph 3 1 1 2 Remove the 4 screws that secure the Transducer 8 to the Base 10 The screws are accessible at the bo...

Страница 19: ...h the balance is rated for a different voltage use an appropriate adapter to match the supply voltage 4 1 2 Segment Display Test 1 Turn the balance on by pressing ON ZERO Off all segments are enabled...

Страница 20: ...g should return to 0g 1 count 4 Repeat this test for ten readings If the standard deviation of the readings is less than 1 count the balance passes the Repeatability Test Adjustment If the deviation f...

Страница 21: ...this test must be adjusted to ASTM Class 4 Tolerance or better NOTE The balance must pass the Off Center Load Test and Re peatabilityTestbeforetheLinearity Testmaybeperformed Test Loads do not have to...

Страница 22: ...hedownstop Tomakeadjustments theTransducermustberemovedfrom the balance Perform procedures 3 1 1 and 3 1 3 and remove the Transducer from the balance NOTE This procedure will require feeler gauges ran...

Страница 23: ...in Max mm in SPx123 0 40 0 016 0 45 0 018 SPx202 0 60 0 024 0 70 0 028 SPx402 0 50 0 020 0 60 0 024 SPx401 0 75 0 030 0 90 0 035 SPx601 0 65 0 026 0 75 0 030 SPx2001 0 65 0 026 0 90 0 035 SPx4001 0 6...

Страница 24: ...4 6 CHAPTER 4 TESTING www mirvesov ru 495 921 44 57 All manuals and user guides at all guides com...

Страница 25: ...nt is made The balance MUST meet the parameters of all applicable specifications in this manual If further technical information is needed in the United States call Ohaus Aftermarket toll free 1 800 5...

Страница 26: ...www mirvesov ru 495 921 44 57 All manuals and user guides at all guides com a l l g u i d e s c o m...

Страница 27: ...t 2000g Scout Pro SP2001 SPX2001 8 71160408 Kit Transducer replacement 6000g Scout Pro SP6001 SPX6001 8 71160407 Kit Transducer replacement 4000g Scout Pro SP4001 SPX4001 8 71160406 Kit Transducer rep...

Страница 28: ...ut Pro All 14 71147994 LCD Scout Pro All 15 71160313 Cal Lock Scout Pro All 16 71160323 Connector Battery Scout Pro All 17 12102033 Weigh Below cover Scout Pro All 18 71160309 Lock sub platform Scout...

Страница 29: ...following URL HTTP fileshare ohaus com docushare a Click on Login b Login ServiceA Password aservice Both are case sensitive c Click on the Xerox logo at the upper left d Click on Technical Support e...

Страница 30: ...5 Install the modified RS232 USB module in the balance being repaired NOTE If using the customers interface do not forget to remove the modification after the repair is complete 6 Plug the interface c...

Страница 31: ...Click on the Port Settings tab 7 Click on the Advanced button 8 Change the COM port number to an unused port 9 Click on OK twice 9 Close the System Properties box 10 Close the Control Panel Window 11...

Страница 32: ...lick on the Test Command tab and test the communications 4 In this example see figure A 6 the P command resulted in the 0 02g being written in the lower screen The V command resulted in Sr 1 10 being...

Страница 33: ...load cell serial number and click on Load cell C in the serial number box 8 The bar G is a graphical indication of the portion of the transducer output that is being used A 6 USING CONFIG TOOL AFTER R...

Страница 34: ...port Baudrate 2400 ByteSize 7 Parity None 9 The balance will respond to the V command but not the others 10 The bar graph will respond to pressure on the platform 11 Enter the balance serial number 1...

Страница 35: ...On Zero Off button 6 Press and release the On Zero Off button The display will show C blinking and then the second mass value 7 Place masses totaling this value on the balance and press the On Zero O...

Страница 36: ...the first mass value 9 Place masses totaling this value on the balance and press the ON ZERO Off button 10 Press and release the ON ZERO Off button The display will show C blinking and then the second...

Страница 37: ...O Off button Thebalancewillreturntotheweighingmode 15 Press and hold the ON ZERO Off button until the display shows OFF 16 Remove power from the balance 17 Remove either the short on U9 or the short o...

Страница 38: ...B 4 APPENDIX B FACTORY CALIBRATIONS www mirvesov ru 495 921 44 57 All manuals and user guides at all guides com...

Страница 39: ...www mirvesov ru 495 921 44 57 All manuals and user guides at all guides com...

Страница 40: ...P N 80250973 SERVICE MANUAL SCOUT PRO ELECTRONIC BALANCES 80250973 All manuals and user guides at all guides com...

Отзывы: