4-5
CHAPTER 4 TESTING
TABLE 4-3. TEST MASS VALUES
Model
Mass 1
Mass 2
SPx202
250
350
SPx402
500
700
SPx401
500
900
SPx601
700
900
SPx2001
3000
7000
SPx4001
5000
7000
SPx6000
7000
9000
TABLE 4-4. TRANSDUCER ADJUSTMENT TOLERANCES
Model
Min(mm/in)
Max(mm/in)
SPx123
0.40/0.016
0.45/0/018
SPx202
0.60/0.024
0.70/0.028
SPx402
0.50/0/020
0.60/0.024
SPx401
0.75/0.030
0.90/0.035
SPx601
0.65/0.026
0.75/0.030
SPx2001
0.65/0.026
0.90/0.035
SPx4001
0.60/0.024
0.70/0.028
SPx6000
0.50/0/020
0.60/0.024
4.3 Down Stop Adjustment (Cont.)
After the transducer has been adjusted, assemble the balance retest and recalibrate.