PXP Series Balance
EN-21
5.3.14 Approved Mode
Use this menu to set the Legal for Trade status.
Off = standard operation.
On = operation complies with Legal Metrology regulations.
Note:
When Approved Mode is set to On, the menu settings are affected as follows:
Calibration Menu:
x
For InCal models, only Internal Calibration is available. All other functions are hidden.
Balance Setup Menu:
x
Filter Level is locked at the current setting.
x
Auto Zero Tracking is limited to 0.5 Division and Off. The selected setting is locked.
x
Auto Tare is locked at current setting.
x
Graduations are forced to 1 Division and the menu item is hidden.
Communication Menu (Communication->Print Settings->Print Output):
x
Stable Weight Only is locked On.
x
Numeric Value Only is locked Off.
Communication Menu (Communication->Print Settings->Auto Print):
x
Auto print mode selections are limited to Off, On Stability, and Interval. Continuous is not available.
Lockout Menu:
x
Menu is hidden
Note:
The security switch located at the rear of the balance must be in the locked position to set Approved Mode to
ON. The security switch must be in the unlocked position to set Approved Mode to Off. See Section 6.
5.4 Weighing Units
Enter this sub-menu to activate the desired units of measure.
PXP balances provide a choice of 21 units, which are all set On by default.
Note:
Due to national laws, the balance may not include some of the units of measure listed.
Display
Unit
Display
Unit
g
Gram
dwt
Pennyweight
kg
Kilogram
mo
Momme
t
Ton
msg
Mesghal
mg
Milligram
tl H
HK tael
ug
Microgram
tl S
SG tael
ct
Carat
tl T
TW tael
N
Newton
tcl
Tical
lb
Pound
tola
Tola
oz
Ounce
baht
Baht
ozt
Troy ounce
lboz
lb:oz
GN
Grain