PXP Series Balance
EN-11
6. Remove the reference sample and place the test
object on the pan. The ratio of the test object to the
reference sample weight is displayed as a
percentage.
7. To view the reference sample weight or the test
object weight, press the
Function / Mode
button.
Item Settings
4.4 Formulation
Note:
Before using any application, be sure the balance has been leveled and calibrated.
Use this application for compounding and recipe making. The maximum number of components is 50.
Formulation
1. Press and hold the
Function / Mode
button until
Formulation
appears.
2. After confirmation by pressing
Yes
, the message
"Place component 1" will appear on the screen.
3. Place the first ingredient on the pan (or in a tared
container) and press the
Function / Mode
button to
store the first weight.
Note:
If the previously established reference sample weight
needs to be kept, press
No
when the message "Clear
reference?" displays.
Printing Setup:
Changing printing setup. See Section 7 for more information.