Error Codes List (Cont.)
Error Codes List (Cont.)
Error Codes List (Cont.)
Error Codes List (Cont.)
Error Codes List (Cont.)
CheckSum Errors
8.2
Power-on load out of specification: Balance was turned on with load on platform
or platform off balance. No load may be on platform when turned on and platform
must be in place.
8.3
Rated capacity exceeded. Remove excessive weight from platform.
8.4
Underload condition on balance. Check that the proper platform and platform
support are installed.
9.0
Bad factory checksum. If error persists, have the balance serviced.
9.5
Bad factory calibration checksum. If error persists, have the balance serviced.
9.6
Bad mode checksum. Turn the balance off using the front panel controls. If
the error persists, have the balance serviced.
9.7
Invalid setup data checksum. Check Setup, User and Print menus (when
RS232 is installed) settings. If possible, try to enter menus and exit using END
to restore menu settings. May be caused by a faulty component, or in rare cases,
a severe static charge. If error persists, balance must be serviced.
9.8
Hardware error causing invalid calibration data checksum. Balance may need
recalibration - particularly linearity calibration. If error persists, balance must be
serviced.
9.9
Invalid temperature compensation checksum. Balance will work with default
temperature compensation data, however, error will occur each time balance is
turned on. Have balance serviced.
SERVICE INFORMATION
SERVICE INFORMATION
SERVICE INFORMATION
SERVICE INFORMATION
SERVICE INFORMATION
If the Troubleshooting section does not resolve or describe your problem, you will need
to contact an authorized Ohaus Service Agent. For Service assistance in the United
States, please call Ohaus Corporation toll-free at (800) 526-0659, an Ohaus Product
Service Specialist will be available to help you.