21
Cal Test (Cont.)
Cal Test (Cont.)
Cal Test (Cont.)
Cal Test (Cont.)
Cal Test (Cont.)
4.
, no mass on the platform.
5.
, value of mass which
must be placed on the platform.
6.
allows the selection of either 25%, 50%, 75%, or 100% of full
span to be used to calibrate the balance.
5. Place required mass on platform.
6.
. While -C- is
displayed, the balance weighs the test mass based on current calibration data,
then displays the difference between the measured value and requested value.
The display shown is an example of when the test mass equals the mass value
stored in memory.
7. Remove mass from platform. After a short period of time, the balance returns
to the weighing mode.
USER MENU
USER MENU
USER MENU
USER MENU
USER MENU
RESET
YES
YES
YES
YES
YES/NO
AL
0, 1
1
1
1
1, 2, 3
STABILITY
.5d
5d
5d
5d
5d, 1d, 2d, 5d
AUTO ZERO
OFF, .5d
.5d
.5d
.5d
.5d, 1d,
3d
BEEP
ON/OFF
OFF
OFF
OFF
OFF
END, User
USER MENU
USER MENU
USER MENU
USER MENU
USER MENU
The User menu is used to adapt the balance to environmental conditions. It
contains submenus which enable you to turn features on or off, and program
balance parameters.
Reset
changes all submenus to original factory default
settings.
Reset
does not appear if menu has been locked out.
AL
specifies the
averaging level.
STB
specifies the desired stability range.
Auto Zero
sets the
automatic zero threshold.
Beep,
when set on, provides audible tones to signify
various balance conditions.
End User
is used to exit the Setup menu and store
the selections. The following figure illustrates the sequence in which submenus
appear on the User menu. Items shown in bold type are the default settings.