1.7. TECHNICAL DATA
1.7.1 Acceleration and Deceleration times FC5706.
Acceleration values
Deceleration values
Rotor-number
level 0
level 9
level 0
level9
30130880 6x5ml
35
8
25
7
30130877 12x5ml
95
33
257
51
30130878 6x50ml
91
41
274
52
in seconds
Acceleration time
Deceleration time
from 0 min
-1
-> U
max
from U
max
-> 0 min
-1
1.7.2 Imbalance shut-off data FC5706 (120V / 230 V )
Rotor-
number
Shut off speed (approx.
values) in rpm
Permitted imbalance in
gram
Imbalance shut-off in gram
30130877
2750
11
12
30130878
2850
10
11
30130880
2850
4
5
1-5